This manual provides operating instructions, maintenance procedures including battery and probe base replacement, error message descriptions, and technical…
The iCare TA01 is a handheld rebound tonometer used for measuring intraocular pressure (IOP) without anesthesia. It utilizes a patented induction-based method with disposable probes to ensure accuracy and prevent cross-contamination during the diagnosis, follow-up, and screening of glaucoma.
This document serves as both a user guide and maintenance manual. It details how to load disposable probes, perform measurements, interpret error messages (such as E01-E07), replace batteries and probe bases, and access stored measurement values. It also includes safety instructions and technical specifications.
| Type | TA01 |
|---|---|
| Dimensions | 13 – 32 mm (W) * 45 – 80 mm (H) * 230 mm (L) |
| Weight | 155 g (without batteries), 250 g (with 4 x AA batteries) |
| Power supply | 4 x AA batteries |
| Measurement range | 1-99 mmHg |
| Accuracy (5-30 mmHg) | ±2.8 mmHg |
The batteries are low and need to be replaced.
Unscrew the collar, tilt the tonometer downward to remove the old base, insert a new one, and screw the collar back in.
The distance should be 4-8 mm from the tip of the probe to the cornea of the eye.
No, measurements should be done without anesthetic because it may reduce the tonometer reading.