HP 8752A Vector Network Analyzer Application Note

This document describes a method to measure oscilloscope input impedance-frequency characteristics using an HP 8752A vector network analyzer.


HP 8752A Vector Network Analyzer Application Note - cover page
Brand
HP
Category
Other Electronics
Document type
Application Note
Language
Chinese
Pages
6
File format
PDF
File size
511 KB
Published
14 October, 2026
Updated
14 October, 2026
MD5 checksum
A1E51A406817EB360CCD6458C49CFAEF

About this manual

This document introduces a new measurement method for determining the impedance-frequency characteristics of an oscilloscope input port. Traditional methods using DC resistance and low-frequency capacitance measurements only provide static data and cannot reflect the high-frequency behavior of the device. The proposed method utilizes an HP 8752A vector network analyzer to measure the complex impedance while the oscilloscope is powered on and excited by a full-bandwidth sine wave signal. This manual is written in Chinese.

What's inside

  • Introduction
  • Impedance Model and Measurement Theory
  • Direct Impedance Measurement
  • Input Capacitance Measurement
  • Measurement Method Comparison
  • Measurement System Setup
  • Calibration Procedure
  • Measurement Execution

Frequently asked questions

What frequency range does the HP8752A cover for this measurement?

The measurement is performed with a signal excitation ranging from 0.3 MHz to 100 MHz.

What is the typical value of the series resistance R in the oscilloscope input circuit?

The series resistance R between the probe lead and the preamplifier is typically 20Ω to 50Ω.

How many times must the trace be averaged to achieve a stable Smith chart display?

You must wait for 16 sweeps, which takes approximately 3 minutes.