This manual provides operating instructions for the Hach DR/820, DR/850, and DR/890 portable colorimeters, covering setup, operation, data storage…
The document is a user manual for the Hach DR/800 series of portable colorimeters, specifically models DR/820, DR/850, and DR/890. It details the instrument's description as an LED-filter photometer controlled by a microprocessor, suitable for laboratory or field use.
The manual covers comprehensive topics including safety warnings, technical specifications, instrument setup (battery installation, date/time setting), operation procedures (zeroing, sample analysis, calibration curves), creating user programs, data storage and retrieval via RS232/IR, maintenance instructions, and a troubleshooting guide for error codes. This manual is written in Chinese.
| Wavelength Range (DR/890) | 420, 520, 560, 610 nm |
|---|---|
| Wavelength Range (DR/850) | 520, 610nm |
| Wavelength Range (DR/820) | 520nm |
| Wavelength Accuracy | ±1 nm |
| Light Source | LED |
| Detector | Silicon Photodiode |
| Display | 4-digit LCD, 1.5cm character height |
| Power Source | 4 AA alkaline batteries |
| Dimensions | 23.6×8.7×4.7 cm (9.3×3.4×1.9 in) |
| Weight | 470 g (1 lb) |
The wavelength accuracy is ±1 nm.
The typical battery life is 6 months.
The instrument operates between 0 to 50°C (32 to 122°F).