Fuller Labs RTG-96K Rickettsia typhi EIA IgG Antibody Kit Instructions for Use

Instructions for the Fuller Labs RTG-96K kit, an indirect enzyme immunoassay for detecting IgG antibodies against Rickettsia typhi in human serum or plasma.


Fuller Labs RTG-96K Rickettsia typhi EIA IgG Antibody Kit Instructions for Use - cover page
Brand
Fuller Labs
Document type
User Manual
Language
English
Pages
3
File format
PDF
File size
180 KB
Published
16 October, 2026
Updated
16 October, 2026
MD5 checksum
BD6E6869C076CFADF6B08500CA221A58

About this manual

The Fuller Labs RTG-96K is an indirect enzyme immunoassay intended for in-vitro diagnostic use to detect and semi-quantitate IgG class human antibody to Rickettsia typhi. The kit utilizes species-specific protein (rOmp B) and lipopolysaccharide antigen purified from Rickettsia typhi coated on EIA microwells.

What's inside

  • Intended Use
  • Summary and Explanation of Test
  • Reagents and Materials Supplied
  • Specimen Collection
  • Preparation of Samples and Reagents
  • Assay Procedure
  • Quality Control and Acceptance Criteria

Specifications

Test Volume96 tests
Storage Temperature2-8˚C
Incubation Time (Step 1)60 minutes
Incubation Time (Step 2)30 minutes
Reaction Time10 minutes
Reading Wavelength450nm

Frequently asked questions

What is the intended use of this kit?

The kit is intended for the detection and semi-quantitation of IgG class human antibody to Rickettsia typhi, to be used as an aid in the diagnosis of human infection by this pathogen.

What are the acceptance criteria for the Negative Control?

The absorbance value must be less than 0.2 units.

How is a positive result determined using the Cutoff Calibrator?

Indices above 1.2 are considered positive, while indices below 0.8 are negative.