FRYKA B35-50 Cold Box Operating Instructions covering erection, technical design, starting, temperature control, alarm unit, service, and parts list.
The FRYKA B35-50 Cold Box is a laboratory cooler with a 35-liter stainless steel chamber and outside dimensions of 585 mm wide, 765 mm deep, and 540 mm high. It must be installed with at least 30 cm of free space on both sides and should stand for at least 12 hours before the cooling machine is started.
The unit uses a fully hermetic compressor with electronic on-off regulation and a digital display, operating from -10°C to -50°C. It includes an alarm unit with temperature and error monitoring, RS485 service connections, and a potential-free alarm contact; maintenance mainly requires checking the air condenser at regular intervals.
| Chamber volume | 35 liters |
|---|---|
| Outside dimensions | 585 x 765 x 540 mm |
| Inside dimensions | 425 x 300 x 285 mm |
| Operating range | -10°C to -50°C |
| Supply voltage | 230 V / 50 Hz |
| Current | 5.0 A |
| Refrigerant | R 507-x, 160 g |
| Control precision | +/- 1 K |
The operating range is -10°C to -50°C.
It should stand at its place for at least 12 hours before starting.
At least 30 cm must remain free at the right and left side.
A dusty air-condenser can lead to overheating of the cooling machine.