Foxboro 875CR COMMUNICATOR Series analyzer measures contacting conductivity and resistivity with dual sensor inputs, dual alarms and 4-20 mA outputs.
The Foxboro 875CR is a microprocessor-based, line-powered intelligent analyzer in the COMMUNICATOR Series, used with Foxboro 871CR or 871CC sensors for either contacting conductivity or resistivity measurement. It is available as a panel mounted analyzer and a field mounted analyzer, the latter carrying NEMA 4X and IEC IP65 protection. Functions include measurement display, dual analog outputs, dual relay contacts and an RS-232 serial port for remote configuration.
The document describes menu-driven configuration with basic and advanced levels, dual sensor input in dual or redundant mode, an RS-232 port with an optional Windows-based configuration utility, and a choice of digital communications PWAs, starting with HART. Resistivity ranges are selectable from 0.2 kΩ •cm through 20 MΩ•cm and conductivity ranges from 0 to 1 µS/cm through 0 to 20 mS/cm, both temperature corrected.
| Measurement | Contacting conductivity or resistivity |
|---|---|
| Resistivity range | 0.2 kΩ•cm through 20 MΩ•cm (temperature corrected) |
| Conductivity range | 0 to 1 µS/cm through 0 to 20 mS/cm (temperature corrected) |
| Resistivity system accuracy | ±1.06% rms |
| Sensor compatibility | Foxboro 871CR and 871CC sensors |
| Analog outputs | Two independent, isolated 4 to 20 mA outputs |
| Alarm relays | Two independent, isolated alarm relays |
| Serial port | Built-in, non-isolated RS-232 |
| Protection | NEMA 4X and IEC IP65 for field mounted analyzer |
| Event history log | Up to 100 events |
| Application configurations | Up to 3 sets stored |
| Standards | USP23/24 compliant |
It is compatible with Foxboro 871CR and 871CC sensors for contacting conductivity or resistivity measurement.
Either one or two contacting type sensors can be connected, and with two sensors the analyzer can be set to dual or redundant mode.
Resistivity ranges are selectable from 0.2 kΩ•cm through 20 MΩ•cm and conductivity ranges from 0 to 1 µS/cm through 0 to 20 mS/cm, both temperature corrected.
At initial product release, Foxboro offers a digital communications PWA for the HART communications protocol.