This is the Simplified Chinese user manual for the Fluke ESA620 Electrical Safety Analyzer, covering safety, operation, and specifications.
This document serves as the user manual for the Fluke ESA620 Electrical Safety Analyzer. It provides instructions on safety precautions, setup, operation, and maintenance procedures for the device. The manual includes a table of contents and detailed sections on connecting the device to power and under test equipment, as well as adjusting display settings and accessing functions. This manual is written in Chinese.
| Device Current Range | 0 – 20 A AC RMS |
|---|---|
| Device Current Accuracy | 读数的 5 % ±(取 2 位数或 0.2 A,以较大值为准) |
| Leakage Current Range | 0.0 至 199.9 µA, 200 至 1999 µA, 2.00 至 10.00 mA |
| Insulation Resistance Range | 0.5 至 20 MΩ, 20 至 100 MΩ |
| Test Voltage Source | 110 % ±5 %, current limited to 7.5 mA ±25 % |
Fluke Biomedical recommends calibrating the analyzer at least once every 12 months.
The user is responsible for shipping the product back to Fluke Biomedical and paying the shipping costs.
The ESA620 Electrical Safety Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.