Fluke ESA620 Electrical Safety Analyzer User Manual

This is the Simplified Chinese user manual for the Fluke ESA620 Electrical Safety Analyzer, covering safety, operation, and specifications.


Fluke ESA620 Electrical Safety Analyzer User Manual - cover page
Brand
Fluke
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
34
File format
PDF
File size
12.3 MB
Published
15 October, 2026
Updated
15 October, 2026
MD5 checksum
A3FC18171C54BED3F766E6152896D4B1

About this manual

This document serves as the user manual for the Fluke ESA620 Electrical Safety Analyzer. It provides instructions on safety precautions, setup, operation, and maintenance procedures for the device. The manual includes a table of contents and detailed sections on connecting the device to power and under test equipment, as well as adjusting display settings and accessing functions. This manual is written in Chinese.

What's inside

  • 简介
  • 安全须知
  • 指定用途
  • 分析仪开箱
  • 熟悉仪器
  • 连接到电源
  • 将被测设备连接至分析仪
  • 启动分析仪
  • 调节显示屏对比度
  • 维护

Specifications

Device Current Range0 – 20 A AC RMS
Device Current Accuracy读数的 5 % ±(取 2 位数或 0.2 A,以较大值为准)
Leakage Current Range0.0 至 199.9 µA, 200 至 1999 µA, 2.00 至 10.00 mA
Insulation Resistance Range0.5 至 20 MΩ, 20 至 100 MΩ
Test Voltage Source110 % ±5 %, current limited to 7.5 mA ±25 %

Frequently asked questions

How often should the analyzer be calibrated?

Fluke Biomedical recommends calibrating the analyzer at least once every 12 months.

Who is responsible for shipping the product for warranty repair?

The user is responsible for shipping the product back to Fluke Biomedical and paying the shipping costs.

Where is the ESA620 manufactured?

The ESA620 Electrical Safety Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.

Download PDF (12.3 MB)