Fluke 771 Calibration Manual Supplement

Web-only supplement (Issue 8) to the Fluke 771 Calibration Manual with eight change notes on adjustment procedures, specifications and maintenance.


Fluke 771 Calibration Manual Supplement - cover page
Brand
Fluke
Category
multimeter
Document type
Service Manual
Language
English
Pages
4
File format
PDF
File size
164 KB
Published
18 November, 2026
Updated
18 November, 2026
Model
771
MD5 checksum
2F8F7080953F2CAA644CC8E45C1A6B8E

About this manual

This is a web-only supplement to the Fluke 771 Calibration Manual, issued as Issue 8 with a print date of November 2007 and a page count of 3. It contains information needed to keep the base manual accurate, presented as eight numbered change notes that replace or add steps in the adjustment procedures and correct specification and maintenance text.

The changes cover the Temperature Adjustment Procedure, High Range Adjustment Procedure, Phase Adjustment Procedure and Calibration Adjustment Procedure, including a replacement Figure 5 in which the clamp direction is reversed. The supplement also changes battery life to 40 hours, the storage temperature range to -25 °C to 60 °C, adds AC noise rejection, measurable conductor diameter and pollution degree information, and adds notes on cleaning the jaw sensor and on what to do after the Jaw Assembly is replaced.

What's inside

  • Temperature Adjustment Procedure
  • High Range Adjustment Procedure
  • Phase Adjustment Procedure
  • Calibration Adjustment Procedure
  • Specifications
  • Maintenance
  • Symbols Table
  • Replaceable Parts

Specifications

Battery Life40 hours
Storage Temperature-25 °C to 60 °C
EMI, RFI, EMCMeets all applicable requirements in EN 61326-1; for field strengths of 1 V/m up to 3 V/m, add 1 mA to specifications
AC Noise RejectionAt 5 Hz, the Meter's reading is the mean value of the measured current
Diameter of Measurable Conductor4.5 mm max
Pollution Degree2
Safety CategoryCAT II 300 V

Frequently asked questions

What is the new battery life specification for the Fluke 771?

The supplement changes the specification from 45 working hours to a battery life of 40 hours.

What is the updated storage temperature range?

The storage temperature is changed to -25 °C to 60 °C.

What must be done if the Jaw Assembly has been replaced?

Complete the Calibration Adjustment Procedure and the Phase Adjustment Procedure before continuing with the remaining adjustment procedures.

How should the sensor surfaces in the clamp jaw be cleaned?

Clean them with a soft cloth and a solution of Isopropyl Alcohol as needed.