Web-only supplement (Issue 8) to the Fluke 771 Calibration Manual with eight change notes on adjustment procedures, specifications and maintenance.
This is a web-only supplement to the Fluke 771 Calibration Manual, issued as Issue 8 with a print date of November 2007 and a page count of 3. It contains information needed to keep the base manual accurate, presented as eight numbered change notes that replace or add steps in the adjustment procedures and correct specification and maintenance text.
The changes cover the Temperature Adjustment Procedure, High Range Adjustment Procedure, Phase Adjustment Procedure and Calibration Adjustment Procedure, including a replacement Figure 5 in which the clamp direction is reversed. The supplement also changes battery life to 40 hours, the storage temperature range to -25 °C to 60 °C, adds AC noise rejection, measurable conductor diameter and pollution degree information, and adds notes on cleaning the jaw sensor and on what to do after the Jaw Assembly is replaced.
| Battery Life | 40 hours |
|---|---|
| Storage Temperature | -25 °C to 60 °C |
| EMI, RFI, EMC | Meets all applicable requirements in EN 61326-1; for field strengths of 1 V/m up to 3 V/m, add 1 mA to specifications |
| AC Noise Rejection | At 5 Hz, the Meter's reading is the mean value of the measured current |
| Diameter of Measurable Conductor | 4.5 mm max |
| Pollution Degree | 2 |
| Safety Category | CAT II 300 V |
The supplement changes the specification from 45 working hours to a battery life of 40 hours.
The storage temperature is changed to -25 °C to 60 °C.
Complete the Calibration Adjustment Procedure and the Phase Adjustment Procedure before continuing with the remaining adjustment procedures.
Clean them with a soft cloth and a solution of Isopropyl Alcohol as needed.