Fluke 754 HART Mode Users Guide covering HART channel selection, transmitter connection, setup, service operations, trims, loop test, and calibration.
The Fluke 754 Documenting Process Calibrator combines communication and calibration functions in one tool. This guide explains how to use the HART communication mode, in which the Product communicates over its serial HART interface with a HART transmitter. It refers readers to the 753/754 Users Manual for safety information, analog mode instructions, specifications, and general data.
The guide covers HART channel selection, connecting to a HART transmitter, supported and generic transmitters, communication operations, setup and service operations, HART calibration, loop test, output trim, sensor trim, and duplicating a transmitter’s basic information. It notes that some HART transmitter commands are available only with supported devices.
| HART channel options | HART Port or mA Jack |
|---|---|
| Warranty period | Three years from date of purchase |
Use Setup mode to select HART Channel, then choose HART Port or mA Jack. The selection is kept in Product memory until changed.
No. A Sensor Trim can only be done on a supported transmitter, while an Output Trim can be done on generic and supported transmitters.
Loop Test sends a command to the transmitter to set its output block to a specified value. It can be used to examine the calibration of the output block or verify an external loop reading device.
Yes. The Product can duplicate Basic Setup data from one transmitter to a different transmitter, including generic and supported transmitters.