Fluke 712B RTD Calibrator User Manual

This manual covers the operation, safety, specifications, and maintenance of the Fluke 712B RTD Calibrator, a handheld instrument for measuring and sourcing…


Fluke 712B RTD Calibrator User Manual - cover page
Brand
Fluke
Category
Industrial Supply
Document type
User Manual
Language
English
Pages
38
File format
PDF
File size
1.9 MB
Published
30 October, 2026
Updated
30 October, 2026
Model
712B
MD5 checksum
12C42CD246F712D0FCC40EBAA04B9551

About this manual

The Fluke 712B RTD Calibrator is a handheld, battery-operated instrument designed to measure and source various Resistance-Temperature Detector (RTD) signals. It features an isolated channel for measuring 4-20 mA current and supports multiple RTD types including Pt100, Pt200, Pt500, Pt1000, Ni120, and Cu variants. The device allows users to simulate RTDs, scale the mA channel to temperature, and perform step or ramp mode operations.

What's inside

  • Safety Information
  • Standard Equipment
  • Input and Output Terminals
  • Keys
  • Display
  • Measure mA Current
  • Measure Temperature
  • Source Temperature Signals
  • Specifications
  • Maintenance

Specifications

DC mA Measurement Range0-24 mA
Ohms Measurement Range0.00 Ω to 4000.0 Ω
Operating Temperature-10 °C to 50 °C
Power Supply4 AA alkaline batteries
Weight524 g
RTD Types SupportedPt100, Pt200, Pt500, Pt1000, Ni120, Cu10, Cu50, Cu100

Frequently asked questions

What is the maximum voltage allowed between terminals?

Do not apply more than 30 V between any two terminals, or between any terminal and earth ground.

How long is the warranty period?

The product is free from defects in material and workmanship for three years from the date of purchase.

What happens if the batteries are not replaced when the low battery indicator shows?

Incorrect measurements may occur, so batteries should be replaced when the indicator shows low.