Fluke 5320A Service Supplement provides updates to ground bond source limits and frequency test limits for the 5320A service manual.
This Fluke 5320A Service Supplement contains changes necessary to ensure the accuracy of the 5320A service manual. It is distributed as an electronic manual on CD-ROM, title 5320A, part number 2634346.
Change #1 replaces Table 3-19 with high test current ground bond source limits. Change #2 updates Table 3-24 frequency test limits to a lower limit of 399.92 Hz and an upper limit of 400.08 Hz.
| Frequency Test Lower Limit | 399.92 Hz |
|---|---|
| Frequency Test Upper Limit | 400.08 Hz |
It contains information necessary to ensure the accuracy of the 5320A service manual.
The lower limit is 399.92 Hz and the upper limit is 400.08 Hz.
It is distributed as an electronic manual on CD-ROM.