This document provides a calibration supplement for the Fluke 365, updating symbols, safety compliance standards, and performance test limits.
This manual supplement, issued in July 2016, contains necessary information to ensure the accuracy of the Fluke 365 calibration manual. It updates the symbols table to include compliance with Australian and South Korean EMC standards and revises the safety compliance section to reflect IEC 61010-1 and IEC 61326-1 standards.
| Measurement Category | Category II and Category IV |
|---|---|
| Safety Standard | IEC 61010-1 Pollution Degree 2 |
| EMC Standard | IEC 61326-1: Portable Electromagnetic Environment |
| FCC Classification | Exempt device per clause 15.103 |
| KCC Classification | Class A Equipment (Industrial Broadcasting & Communication Equipment) |
| AC Volts Limit (30V @ 50Hz) | 29.1 V to 30.9 V |
| DC Volts Limit (30V) | 29.2 V to 30.8 V |
| Ohms Limit (300) | 296.5 to 303.5 |
| AC Amps Limit (0.5A @ 50Hz) | 9.3 A to 10.7 A |
It conforms to relevant Australian and South Korean EMC standards.
No, it is intended for business environments and may not provide adequate protection to radio reception in residential environments.
Measurement Category II is applicable to test and measuring circuits connected directly to utilization points.