Fluke 28 II Ex Digital Multimeter Calibration Information

Calibration information for the Fluke 28 II Ex digital multimeter, including safety, specifications, fuse test, performance tests, and calibration adjustment.


Fluke 28 II Ex Digital Multimeter Calibration Information - cover page
Brand
Fluke
Category
multimeter
Document type
Service Manual
Language
English
Pages
20
File format
PDF
File size
882 KB
Published
24 November, 2026
Updated
24 November, 2026
MD5 checksum
03FB4DB8A5600EB8A80C16C205707217

About this manual

Fluke 28 II Ex Digital Multimeter Calibration Information is a technical document for the Fluke 28 II Ex digital multimeter. It covers safety information, international electrical symbols, specifications, general maintenance, fuse test, performance tests, calibration adjustment procedure, and limited lifetime warranty.

The document states that operating instructions are in the 28 II Ex Users Manual and provides Fluke contact information. Its specifications include a 1000 V rms maximum voltage between any terminal and earth ground, a 6000-count display, 3 AAA alkaline batteries, and an IP rating of 67 (non-operating).

What's inside

  • Safety Information
  • International Electrical Symbols
  • General Specifications
  • Detailed Specifications
  • General Maintenance
  • Fuse Test
  • Performance Tests
  • Calibration Adjustment Procedure
  • Limited Lifetime Warranty
  • How to Contact Fluke

Specifications

Maximum voltage between any terminal and earth ground1000 V rms
Fuse for mA inputs440 mA, 1000 V FAST Fuse
Fuse for A inputs11 A, 1000 V FAST Fuse
Display6000 counts, updates 4/sec (19,999 counts in high-resolution mode)
Altitude, operating2,000 meters
Altitude, storage10,000 meters
Operating temperature-15 °C to 50 °C
Storage temperature (without battery)-40 °C to +85 °C
Storage temperature (with battery)-40 °C to +60 °C
Temperature coefficient0.05 X (specified accuracy) / °C (<18 °C or >28 °C)
Electromagnetic compatibilityIn an RF field of 3 V/M, accuracy = specified accuracy +20 counts, except 600 μA dc range total accuracy = specified accuracy +60 counts
Relative humidity0 % to 80 % (0 °C to 35 °C); 0 % to 70 % (35 °C to 50 °C)
Battery type3 AAA Alkaline batteries, NEDA 24A IEC LR03
Approved batteriesDuracell Procell MN2400 LR03; Duracell Plus MN2400 LR03; Varta Max Tech No. 4703; Varta Industrial Alcaline No. 4003; Eveready Energizer No. E92; Rayovac Alkaline AAA (U.S. Type); Panasonic LR03XWA
Battery life400 hrs typical without backlight (Alkaline)
VibrationPer MIL-PRF-28800 for a Class 2 instrument
Shock1 Meter drop per IEC 61010 (3 Meter drop with holster)
Size (H x W x L)4.57 cm x 10.0 cm x 21.33 cm (1.80 in x 3.95 in x 8.40 in)
Size with holster6.35 cm x 10.0 cm x 19.81 cm (2.50 in x 3.95 in x 7.80 in)
Weight567.8 g (1.25 lb)
Weight with holster and Flex-Stand769.8 g (1.70 lb)
Safety complianceComplies with ANSI/ISA S82.01-2004, CAN/CSA C22.2 61010-1-04 to 600 V Measurement Category IV. Licensed by TÜV to EN61010-1, Pollution degree 2
CertificationsCSA, TÜV, P, ATEX, IECEx
IP rating67 (Non-operating. Protected against dust and the effect of immersion up to 1 m for 30 min.)
AC voltage 600.0 mV range accuracy (45-65 Hz)±(0.7 % + 4)
DC voltage 6.000 V range accuracy±(0.05 % + 1)
Resistance 600.0 Ω range accuracy±(0.2 % + 2)
Temperature range-200 °C to +1090 °C (-328 °F to +1994 °F)
Temperature accuracy±(1.0 % + 10) °C; ±(1.0 % + 18) °F
AC current 600.0 μA range accuracy (45 Hz-2 kHz)±(1.0 % + 2)
DC current 600.0 μA range accuracy±(0.2 % + 4)
Capacitance 10.00 nF range accuracy±(1.0 % + 2)
Diode 2.000 V range accuracy±(2.0 % + 1)
Frequency 199.99 Hz range accuracy±(0.005 % + 1)

Frequently asked questions

What is the maximum voltage between any terminal and earth ground?

The maximum voltage between any terminal and earth ground is 1000 V rms.

What batteries does the Fluke 28 II Ex use?

It uses 3 AAA alkaline batteries (NEDA 24A IEC LR03).

What is the IP rating of the meter?

The IP rating is 67 (non-operating), protecting against dust and immersion up to 1 m for 30 min.

Where are operating instructions for the meter found?

Operating instructions are in the 28 II Ex Users Manual.

Download PDF (882 KB)

More Fluke Manuals