Calibration information for the Fluke 28 II Ex digital multimeter, including safety, specifications, fuse test, performance tests, and calibration adjustment.
Fluke 28 II Ex Digital Multimeter Calibration Information is a technical document for the Fluke 28 II Ex digital multimeter. It covers safety information, international electrical symbols, specifications, general maintenance, fuse test, performance tests, calibration adjustment procedure, and limited lifetime warranty.
The document states that operating instructions are in the 28 II Ex Users Manual and provides Fluke contact information. Its specifications include a 1000 V rms maximum voltage between any terminal and earth ground, a 6000-count display, 3 AAA alkaline batteries, and an IP rating of 67 (non-operating).
| Maximum voltage between any terminal and earth ground | 1000 V rms |
|---|---|
| Fuse for mA inputs | 440 mA, 1000 V FAST Fuse |
| Fuse for A inputs | 11 A, 1000 V FAST Fuse |
| Display | 6000 counts, updates 4/sec (19,999 counts in high-resolution mode) |
| Altitude, operating | 2,000 meters |
| Altitude, storage | 10,000 meters |
| Operating temperature | -15 °C to 50 °C |
| Storage temperature (without battery) | -40 °C to +85 °C |
| Storage temperature (with battery) | -40 °C to +60 °C |
| Temperature coefficient | 0.05 X (specified accuracy) / °C (<18 °C or >28 °C) |
| Electromagnetic compatibility | In an RF field of 3 V/M, accuracy = specified accuracy +20 counts, except 600 μA dc range total accuracy = specified accuracy +60 counts |
| Relative humidity | 0 % to 80 % (0 °C to 35 °C); 0 % to 70 % (35 °C to 50 °C) |
| Battery type | 3 AAA Alkaline batteries, NEDA 24A IEC LR03 |
| Approved batteries | Duracell Procell MN2400 LR03; Duracell Plus MN2400 LR03; Varta Max Tech No. 4703; Varta Industrial Alcaline No. 4003; Eveready Energizer No. E92; Rayovac Alkaline AAA (U.S. Type); Panasonic LR03XWA |
| Battery life | 400 hrs typical without backlight (Alkaline) |
| Vibration | Per MIL-PRF-28800 for a Class 2 instrument |
| Shock | 1 Meter drop per IEC 61010 (3 Meter drop with holster) |
| Size (H x W x L) | 4.57 cm x 10.0 cm x 21.33 cm (1.80 in x 3.95 in x 8.40 in) |
| Size with holster | 6.35 cm x 10.0 cm x 19.81 cm (2.50 in x 3.95 in x 7.80 in) |
| Weight | 567.8 g (1.25 lb) |
| Weight with holster and Flex-Stand | 769.8 g (1.70 lb) |
| Safety compliance | Complies with ANSI/ISA S82.01-2004, CAN/CSA C22.2 61010-1-04 to 600 V Measurement Category IV. Licensed by TÜV to EN61010-1, Pollution degree 2 |
| Certifications | CSA, TÜV, P, ATEX, IECEx |
| IP rating | 67 (Non-operating. Protected against dust and the effect of immersion up to 1 m for 30 min.) |
| AC voltage 600.0 mV range accuracy (45-65 Hz) | ±(0.7 % + 4) |
| DC voltage 6.000 V range accuracy | ±(0.05 % + 1) |
| Resistance 600.0 Ω range accuracy | ±(0.2 % + 2) |
| Temperature range | -200 °C to +1090 °C (-328 °F to +1994 °F) |
| Temperature accuracy | ±(1.0 % + 10) °C; ±(1.0 % + 18) °F |
| AC current 600.0 μA range accuracy (45 Hz-2 kHz) | ±(1.0 % + 2) |
| DC current 600.0 μA range accuracy | ±(0.2 % + 4) |
| Capacitance 10.00 nF range accuracy | ±(1.0 % + 2) |
| Diode 2.000 V range accuracy | ±(2.0 % + 1) |
| Frequency 199.99 Hz range accuracy | ±(0.005 % + 1) |
The maximum voltage between any terminal and earth ground is 1000 V rms.
It uses 3 AAA alkaline batteries (NEDA 24A IEC LR03).
The IP rating is 67 (non-operating), protecting against dust and immersion up to 1 m for 30 min.
Operating instructions are in the 28 II Ex Users Manual.