Fluke 187 & 189 Series True RMS Digital Multimeter Service Manual

Service manual for the Fluke 187 and 189 true RMS digital multimeters, covering common faults and repair methods.


Fluke 187 & 189 Series True RMS Digital Multimeter Service Manual - cover page
Brand
Fluke
Category
multimeter
Document type
Service Manual
Language
Chinese
Pages
28
File format
PDF
File size
1.6 MB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
5CFCD8E369C2BA7DBE507A70A2A8D39A

About this manual

This document is a service manual for the Fluke 187 and 189 true RMS digital multimeters. It describes the differences between the two models, mainboard component layout, and repair guidance for common faults.

Covered repairs include test-lead jack faults, moisture-related PCB leakage, UCAL calibration data errors, LCD segment problems, keypad failures, Fluke 189 farad capacitor issues, and current-range faults. This manual is written in Chinese.

What's inside

  • Differences between Fluke 187 and 189
  • Mainboard component layout
  • Common fault: test-lead jack
  • Common fault: moisture leakage
  • Common fault: UCAL calibration data error
  • Common fault: LCD missing or extra segments
  • Common fault: keypad failure
  • Common fault: Fluke 189 farad capacitor
  • Current range repair
  • Chip functions

Specifications

Maximum count50000 counts
DisplayDual display
Measurement functionsAC+DC, dB, nS conductance, diode, capacitance, temperature
AC voltage/current frequency response100 kHz

Frequently asked questions

What can cause a leads error on the Fluke 187/189?

It can be caused by shorted metal contacts in the test-lead jack, a dirty or moist jack, a damaged fuse, or damaged current-range detection resistors.

How is moisture-related leakage on the mainboard repaired?

Remove the mainboard and soak it completely in 95% alcohol for at least 24 hours, then dry it.

What does UCAL on startup indicate?

It indicates a calibration data error; recalibrate the meter or rewrite the AT93C56 EEPROM data.

Why can a Fluke 189 lose its time setting after power is removed?

The farad capacitor may be rusted or leaking; it can be replaced or removed.