Service manual for the Fluke 187 and 189 true RMS digital multimeters, covering common faults and repair methods.
This document is a service manual for the Fluke 187 and 189 true RMS digital multimeters. It describes the differences between the two models, mainboard component layout, and repair guidance for common faults.
Covered repairs include test-lead jack faults, moisture-related PCB leakage, UCAL calibration data errors, LCD segment problems, keypad failures, Fluke 189 farad capacitor issues, and current-range faults. This manual is written in Chinese.
| Maximum count | 50000 counts |
|---|---|
| Display | Dual display |
| Measurement functions | AC+DC, dB, nS conductance, diode, capacitance, temperature |
| AC voltage/current frequency response | 100 kHz |
It can be caused by shorted metal contacts in the test-lead jack, a dirty or moist jack, a damaged fuse, or damaged current-range detection resistors.
Remove the mainboard and soak it completely in 95% alcohol for at least 24 hours, then dry it.
It indicates a calibration data error; recalibrate the meter or rewrite the AT93C56 EEPROM data.
The farad capacitor may be rusted or leaking; it can be replaced or removed.