Fluke 1750 Power Recorder Calibration Manual

Calibration manual for the Fluke 1750 Power Recorder, covering safety, specifications, verification tests, calibration procedures, required equipment, and…


Fluke 1750 Power Recorder Calibration Manual - cover page
Brand
Fluke
Category
Monitoring Equipment
Document type
Service Manual
Language
English
Pages
28
File format
PDF
File size
1.2 MB
Published
10 November, 2026
Updated
10 November, 2026
Model
1750
MD5 checksum
1B8ED6CA66DDDBF66CC72AC0424F63F2

About this manual

Fluke 1750 Power Recorder Calibration Manual is for qualified personnel and covers safety information, specifications, basic maintenance, calibration procedures, verification procedures, and replaceable parts and accessories. It states that the recorder complies with EN 61010-1 600 V CAT IV, 1000 V CAT III overvoltage rating and warns against performing calibration unless qualified.

The manual lists required equipment and test cables, then describes verification tests for AC voltage, neutral voltage, Rogowski and CT current, watt, and impulse, followed by calibration adjustment steps. Its specifications include power, frequency, power factor, harmonic and inter-harmonic analysis, and LAN and wireless interface details.

What's inside

  • Introduction
  • Safety Information
  • Specifications
  • Cleaning and Maintenance
  • Required Equipment
  • Verification Tests
  • Calibration
  • Test Cables

Specifications

Power measurement methodCalculated per IEEE1459 for best performance when distortions exist
Power measurement typeTrue rms calculated continuously: every cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards
Power measurement accuracy± 0.2 % reading ± 0.1 % full scale + current sensor accuracy
Frequency measurement range42.5 to 69 Hz
Frequency measurement accuracy± 10 mHz (10 to 110 % of range, with sine wave)
Reactive power accuracy± 0.2 % reading ± 0.1 % full scale + current sensor accuracy
Power factor measurement range-1.000 (leading) to 0.000 to +1.000 (lagging)
Power factor measurement accuracy± 1 digit from the calculation of each measured value (± 3 digits for total)
Displacement power factor measurement methodCalculated from the phase difference between voltage fundamental and current fundamental
Displacement power factor measurement range- 1.000 (leading) to 0.000 to + 1.000 (lagging)
Displacement power factor measurement accuracy± 0.5 % reading ± 2 % full scale ± 1 digit
Voltage unbalance and phase sequence measurement methodPositive sequence voltage divided by negative sequence voltage, per IEC 61000-4-30
Harmonic voltage and current analysis windowrectangular
Harmonic analysis order1st to 50th order
Harmonic measurement accuracyVoltage / Current: 1st to 20th orders: ± 0.5 % reading ± 0.2 % full scale; 21st to 50th orders: ± 1 % reading ± 0.3 % full scale
Harmonic measurement methodIEC 61000-4-7
Inter-harmonic analysis windowrectangular
Inter-harmonic analysis orders0.5 to 49.5th order
LAN interface connectorRJ-45
LAN speed and type10/100 Base-T, auto MDIX
LAN communications protocolTCP/IP over Ethernet
Wireless controller interface connectionwireless (2.4 GHz radio)
Wireless speedup to 700 kbit/second
Wireless communications protocolBluetooth SPP
Voltage dip measurement typeTrue rms (one cycle calculation by overlapping each half cycle)
Voltage dip displayed dataAmplitude and duration of dip or interruption
Voltage dip measurement accuracySame as rms voltage
Voltage dropout measurement typeSame as Voltage Dip

Frequently asked questions

What is the warranty period for the Fluke 1750?

The warranty period is one year from the date of shipment, and parts, product repairs, and services are warranted for 90 days.

What frequency range does the Fluke 1750 measure?

It measures frequency from 42.5 to 69 Hz.

What LAN connector does the Fluke 1750 use?

It uses an RJ-45 LAN interface connector.

What wireless protocol does the Fluke 1750 use?

It uses Bluetooth SPP over a wireless 2.4 GHz radio interface.