Calibration manual for the Fluke 1750 Power Recorder, covering safety, specifications, verification tests, calibration procedures, required equipment, and…
Fluke 1750 Power Recorder Calibration Manual is for qualified personnel and covers safety information, specifications, basic maintenance, calibration procedures, verification procedures, and replaceable parts and accessories. It states that the recorder complies with EN 61010-1 600 V CAT IV, 1000 V CAT III overvoltage rating and warns against performing calibration unless qualified.
The manual lists required equipment and test cables, then describes verification tests for AC voltage, neutral voltage, Rogowski and CT current, watt, and impulse, followed by calibration adjustment steps. Its specifications include power, frequency, power factor, harmonic and inter-harmonic analysis, and LAN and wireless interface details.
| Power measurement method | Calculated per IEEE1459 for best performance when distortions exist |
|---|---|
| Power measurement type | True rms calculated continuously: every cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards |
| Power measurement accuracy | ± 0.2 % reading ± 0.1 % full scale + current sensor accuracy |
| Frequency measurement range | 42.5 to 69 Hz |
| Frequency measurement accuracy | ± 10 mHz (10 to 110 % of range, with sine wave) |
| Reactive power accuracy | ± 0.2 % reading ± 0.1 % full scale + current sensor accuracy |
| Power factor measurement range | -1.000 (leading) to 0.000 to +1.000 (lagging) |
| Power factor measurement accuracy | ± 1 digit from the calculation of each measured value (± 3 digits for total) |
| Displacement power factor measurement method | Calculated from the phase difference between voltage fundamental and current fundamental |
| Displacement power factor measurement range | - 1.000 (leading) to 0.000 to + 1.000 (lagging) |
| Displacement power factor measurement accuracy | ± 0.5 % reading ± 2 % full scale ± 1 digit |
| Voltage unbalance and phase sequence measurement method | Positive sequence voltage divided by negative sequence voltage, per IEC 61000-4-30 |
| Harmonic voltage and current analysis window | rectangular |
| Harmonic analysis order | 1st to 50th order |
| Harmonic measurement accuracy | Voltage / Current: 1st to 20th orders: ± 0.5 % reading ± 0.2 % full scale; 21st to 50th orders: ± 1 % reading ± 0.3 % full scale |
| Harmonic measurement method | IEC 61000-4-7 |
| Inter-harmonic analysis window | rectangular |
| Inter-harmonic analysis orders | 0.5 to 49.5th order |
| LAN interface connector | RJ-45 |
| LAN speed and type | 10/100 Base-T, auto MDIX |
| LAN communications protocol | TCP/IP over Ethernet |
| Wireless controller interface connection | wireless (2.4 GHz radio) |
| Wireless speed | up to 700 kbit/second |
| Wireless communications protocol | Bluetooth SPP |
| Voltage dip measurement type | True rms (one cycle calculation by overlapping each half cycle) |
| Voltage dip displayed data | Amplitude and duration of dip or interruption |
| Voltage dip measurement accuracy | Same as rms voltage |
| Voltage dropout measurement type | Same as Voltage Dip |
The warranty period is one year from the date of shipment, and parts, product repairs, and services are warranted for 90 days.
It measures frequency from 42.5 to 69 Hz.
It uses an RJ-45 LAN interface connector.
It uses Bluetooth SPP over a wireless 2.4 GHz radio interface.