Supplement to the Fluke 1550C/1555 calibration manual with symbol, specification, and voltage test table changes.
This supplement contains information necessary to ensure the accuracy of the Fluke 1550C/1555 calibration manual. It adds symbols to the Symbols table and revises the General Specifications section.
It also replaces Table 6 with a voltage measurement test table listing two test steps and their UUT display limits.
| IP Rating | IEC 60529: IP40 |
|---|---|
| Safety | IEC 61010-1: 600 V CAT IV / 1000 V CAT III, Pollution Degree 2 |
| Electromagnetic Environment | IEC 61326-1: Portable |
| Electromagnetic Compatibility | Class A Equipment (Industrial Broadcasting & Communication Equipment); applies to use in Korea only |
| Voltage Measurement Test Step 1 UUT Display Limits | 30 V to 46 V |
| Voltage Measurement Test Step 2 UUT Display Limits | 202 V to 278 V |
The supplement adds a symbol for conformity to relevant South Korean EMC Standards and measurement category definitions for CAT II, CAT III, and CAT IV.
The specified IP rating is IP40 according to IEC 60529.
The specified safety rating is IEC 61010-1: 600 V CAT IV / 1000 V CAT III, Pollution Degree 2.
The minimum is 30 V and the maximum is 46 V.