Fluke 1550C/1555 Calibration Manual Supplement

Supplement to the Fluke 1550C/1555 calibration manual with symbol, specification, and voltage test table changes.


Fluke 1550C/1555 Calibration Manual Supplement - cover page
Brand
Fluke
Category
Other Electronics
Document type
Other
Language
English
Pages
2
File format
PDF
File size
53 KB
Published
10 November, 2026
Updated
10 November, 2026
MD5 checksum
185AB1526BF3DBE9263D49A63CA74E05

About this manual

This supplement contains information necessary to ensure the accuracy of the Fluke 1550C/1555 calibration manual. It adds symbols to the Symbols table and revises the General Specifications section.

It also replaces Table 6 with a voltage measurement test table listing two test steps and their UUT display limits.

What's inside

  • Manual Supplement
  • Change #1
  • Change #2
  • Symbols table
  • General Specifications
  • Voltage Measurement Test

Specifications

IP RatingIEC 60529: IP40
SafetyIEC 61010-1: 600 V CAT IV / 1000 V CAT III, Pollution Degree 2
Electromagnetic EnvironmentIEC 61326-1: Portable
Electromagnetic CompatibilityClass A Equipment (Industrial Broadcasting & Communication Equipment); applies to use in Korea only
Voltage Measurement Test Step 1 UUT Display Limits30 V to 46 V
Voltage Measurement Test Step 2 UUT Display Limits202 V to 278 V

Frequently asked questions

What is added to the Symbols table?

The supplement adds a symbol for conformity to relevant South Korean EMC Standards and measurement category definitions for CAT II, CAT III, and CAT IV.

What IP rating is specified?

The specified IP rating is IP40 according to IEC 60529.

What safety rating is specified?

The specified safety rating is IEC 61010-1: 600 V CAT IV / 1000 V CAT III, Pollution Degree 2.

What are the UUT display limits for voltage measurement test step 1?

The minimum is 30 V and the maximum is 46 V.