Operator's manual with safety, installation, specifications, operation, and troubleshooting for Fisher Scientific tabletop ultrasonic cleaners (FS-20 to…
This manual provides essential safety precautions and proper handling instructions for Fisher Scientific tabletop ultrasonic cleaners. It covers grounding, use of water‑based solutions, and warnings against flammable liquids and mineral acids to protect both the user and the equipment.
The document details equipment specifications, how ultrasonic cleaning works, step‑by‑step operating procedures, cleaning solution recommendations, and troubleshooting tips. Service center locations and warranty information are also included for maintenance and repair support.
| Tank Capacity | 3/4 gal. (2.8 L) |
|---|---|
| Tank Size | L6" W: 5.5" D:6" |
| Overall Size | L: 8.9" W: 8.9" H: 11.3" |
| Weight | 11 lbs. (5KG) |
| Ultrasonic Power | 130W |
| Heater Power | 63W |
| Operating Frequency | 40kHz |
| Voltage Options | 120V or 220V ± 10%, 50/60Hz |
Do not place parts or containers directly on the bottom of the cleaning tank; use a tray or wire to suspend items, keep the unit grounded, unplug before filling, and use only water‑based solutions.
Perform the Glass Slide Test: immerse a frosted microscope slide with a pencil‑drawn "X" in the solution; the ultrasonic action should remove all lead within about 10 seconds.