Instruction manual for the Eutech/Oakton CyberScan DO 110 hand-held dissolved oxygen meter, covering setup, calibration, measurement, memory, printing, and…
The DO 110 is a microprocessor-based hand-held meter for measuring dissolved oxygen in mg/L or ppm, percent saturation, and temperature. It has a water-resistant membrane keypad and a dual LCD display.
This manual covers preparation, calibration, measurement, memory and print functions, advanced setup, CyberComm Portable DAS, probe care and replacement, error messages, troubleshooting, accessories, specifications, warranty, and return procedures.
| Dissolved oxygen range | 0.00 – 19.99 mg/L or ppm |
|---|---|
| Dissolved oxygen resolution | 0.01 mg/L; 0.01 ppm |
| Dissolved oxygen relative accuracy | ± 1.5% of Full Scale |
| Oxygen saturation range | 0.0 – 199.9 % |
| Oxygen saturation relative accuracy | ± 1.5% of Full Scale |
| Temperature range (meter) | -10 to 110 °C (14 to 230 °F) |
| Temperature resolution | 0.1 °C / 0.1 °F |
| Temperature relative accuracy | +/- 0.5 °C (+/- 0.9 °F) |
| Salinity correction range | 0.0 – 50.0 ppt |
| Barometric pressure correction range | 500 to 1499 mm Hg or 66.6 to 199.9 kPA |
| Probe type | Galvanic probe |
| Probe temperature operating range | 0.0 – 50.0 °C |
| Probe response time | 40 seconds to achieve 93% of the reading |
| Number of calibration points | Up to 2 points (0% and 100%) |
| Data logging | 100 |
| RS232 | Yes |
| HOLD function | Yes |
| Auto-off function | Selectable; 20 minutes after last key press |
| Averaging/stability function | Selectable |
| Display | Dual LCD |
| Power requirements | Four 1.5 V AAA-sized batteries (included) or 100/240 VAC SMPS power adapter to 9 VDC, 6W (optional) |
| Battery life | > 700 hours |
| Temperature compensation | Automatic from 0 to 50 °C |
| Operating range | 0 to 50 °C |
| Dimensions (meter) | 19.1 cm (L) x 8.9 cm (W) x 4.5 cm (H) |
| Shipping weight | 920 g |
It measures dissolved oxygen in mg/L or ppm, percent saturation, and temperature.
Up to two points: 0% and 100%.
More than 700 hours.
Yes, the PRINT function can print current measurement or stored data to a printer or computer.