Official service manual for Esaote MyLab ultrasound probes and consumables, covering handling, reprocessing, storage, and safety guidelines.
This service manual provides detailed instructions for handling, storing, inspecting, and reprocessing Esaote MyLab ultrasound probes and needle guide kits. It includes safety precautions, cleaning and sterilization procedures, and guidelines for different probe types including transesophageal, laparoscopic, and intraoperative probes. The document also covers end-of-life disposal and compliance with medical device regulations. It is intended for technicians and medical staff who maintain or operate MyLab ultrasound systems.
| Product Life Time | 7 years from purchase date |
|---|---|
| Maintainability Time | 7 years from purchase date |
| Sterilization Pressure | 2.1 bar |
| Sterilization Time | 7 minutes |
The safety and efficiency of MyLab probes are guaranteed for at least seven (7) years from the purchase date, provided they are used according to instructions.
Do not use the probe until it has been inspected by Esaote Service personnel. Check for electrical damage by measuring leakage current.
Inspect non-invasive probes monthly or after any impact. Inspect endocavity, transesophageal, intraoperative, and laparoscopic probes before every examination.
Steam sterilization requires 2.1 bar pressure, 134°C temperature, and 7 minutes duration.