Instruction manual for the Enraf 854 XTG level gauge, covering commissioning, maintenance, and troubleshooting for technicians.
The Enraf 854 XTG (Xmission Technology Gauge) is a device designed to measure liquid levels and can be programmed to measure two additional interface levels. It utilizes a displacer suspended from a measuring wire to detect variations in buoyancy, providing stable and accurate averaged level measurements.
This manual is intended for technicians involved in the commissioning and service of the instrument. It provides detailed procedures for installation of the measuring drum and displacer, programming, level calibration, and preventive maintenance.
| Vertical movement resolution | 0.05 mm |
|---|---|
| Stepper motor movement | 10 mm per revolution |
| Explosion proof certification | Zone 0 |
The measurement is based on detecting variations in the buoyancy of a displacer suspended from a measuring wire.
It is achieved by sending an interface command that moves the displacer to a position where its apparent weight matches a pre-programmed set point.