This manual provides instructions for the installation, operation, maintenance, and safety precautions for the EliTech D5111 Diode Array Detector.
The EliTech D5111 Diode Array Detector is a module designed for use with high-performance liquid chromatography (HPLC) systems. This user manual details the instrument's composition, installation procedures, operational steps, maintenance requirements, and associated safety precautions to ensure correct usage and equipment longevity. It covers topics ranging from unpacking and stacking to software operation and troubleshooting. This manual is written in Chinese.
| Light Source | Deuterium lamp, Tungsten lamp |
|---|---|
| Response Time | 0s, 1s, 2s |
| Sampling Frequency | 10Hz (max 80Hz) |
| Wavelength Range | 190-800nm |
| Wavelength Repeatability | ≤0.1nm |
| Wavelength Accuracy | ±1nm |
| Spectral Resolution | 1.2nm |
| Array Number | 512 |
| Noise | ≤±1.0×10-5AU |
| Baseline Drift | ≤2×10-4AU/h |
| Linear Range | ≥2.0AU(5%) |
| Detection Cell Pressure Resistance | ≥8MPa |
| Detection Cell Path Length | 10mm |
| Communication | UDP |
| Weight | 16.5Kg |
| Dimensions | 540×400×160 (mm) |
| Operating Voltage | 220V±10% |
| Rated Power | 80W |
Solvents such as concentrated sulfuric acid, concentrated nitric acid, dichloroacetic acid, dichloromethane, chloroform, dimethyl sulfoxide, acetone, and tetrahydrofuran should not be used.
Daily maintenance does not require opening the main cover; repairs requiring cover removal should be delegated to authorized agents or customer service staff.