User manual for Dynasonics TFXP transit time ultrasonic flow meter, covering setup, battery charging, transducer mounting, and operation.
This manual provides step-by-step instructions for operating and maintaining the Dynasonics TFXP transit time ultrasonic flow meter. It covers initial setup, battery charging, transducer mounting methods (W-, V-, Z-mount), pipe preparation, and power connections. The device uses clamp-on transducers to measure fluid velocity without contact, suitable for pipe sizes from 1 to 100 inches. It includes safety features, non-volatile memory, and optional outputs like 4-20mA and RS485. The manual also details installation guidelines, signal strength verification, and troubleshooting.
The TFXP is designed for non-fouling, wear-free operation in various liquids including water, chemicals, and sewage. It supports extended operation with AC power or battery, and includes a data logger and password-protected settings. The device is rated for up to 300°F (150°C) and can be used in harsh environments with proper mounting and protection.
| Battery Type | Lead Acid Gel Cell |
|---|---|
| Battery Life | Up to 24 hours continuous operation |
| Operating Temperature | -40 to 185 °F [-40 to 85°C] |
| Pipe Size Range | 1 - 100 inch [25 - 2540 mm] |
| Maximum Temperature Rating | 300 °F [150°C] |
| Signal Strength Minimum | 2.0% |
The TFXP can operate up to 300°F (150°C), though higher temperatures require consultation with Dynasonics.
Signal strength must be greater than 2.0% for proper operation.