The Intelli Nuclease ELISA Kit (96T per kit) from Shanghai Duoning Biotechnology is intended for quantitative detection of omnipotent nuclease in samples. It uses a double-antibody sandwich principle: antibody-coated microplate wells capture the nuclease, an HRP-labeled antibody completes the complex, and TMB color development measured at 450 nm is proportional to the nuclease concentration, which is calculated from a calibration curve.

This bilingual (Chinese/English) protocol booklet covers the intended purpose, required instruments, kit composition, reagent preparation, the step-by-step assay procedure, interpretation of results, test performance, storage conditions, and cautions such as avoiding sodium azide in samples. This manual is written in Chinese.

What's inside

  • Intended Purpose
  • Principle of Test
  • Instruments Required
  • Composition of kit
  • Preparation of Reagents
  • Assay Procedures
  • Interpretation of Test Results
  • Test Performance
  • Cautions
  • Storage Condition and Stability

Specifications

Specification96T/kit
Limit of detection (LOD)0.06 ng/mL
Lower limit of detection (LLOD)0.2 ng/mL
Linearity range0.2~20 ng/mL
PrecisionIntra-run CV ≤10%, inter-run CV ≤15%
Recovery80%~120%
Storage and stability2~8°C for 12 months; use within 6 weeks after unpacking
Readout wavelength450 nm (dual wavelength 450 nm/650 nm recommended)

Frequently asked questions

What is the sensitivity of the Intelli Nuclease ELISA Kit?

The kit has an LOD of 0.06 ng/mL, a lower limit of detection of 0.2 ng/mL, and a linear range of 0.2–20 ng/mL.

How should the kit be stored?

Store at 2~8°C for 12 months and use it up within 6 weeks after unpacking.

At what wavelength should the microplate be read?

Read the OD at 450 nm within 15 minutes of adding stop solution; dual-wavelength detection at 450 nm/650 nm is recommended.

Can the kit detect omnipotent nuclease from other manufacturers?

Yes, it can also detect nuclease from other manufacturers, but a calibration curve should be established with the corresponding product.