User manual for Daviteq MBRTU-SAL salinity sensor with Modbus RTU output, covering specs, wiring, maintenance, and protocol details.
The Daviteq MBRTU-SAL is a salinity sensor using non-contact electromagnetic measurement. It features automatic temperature compensation via a built-in PT1000 sensor and is suitable for marine, industrial wastewater, pharmaceutical, and biotech applications. The sensor outputs Modbus RTU over RS-485 and includes calibration registers for temperature, zero point, and slope. Maintenance is minimal, with cleaning recommended every 30 days. The sensor is IP68 rated and can operate at depths up to 20 meters.
The manual details wiring, maintenance procedures, Modbus RTU protocol specifications (including baud rate, register addresses, and command examples), and calibration instructions for temperature and salinity. It also includes dimensions, contact information, and references to related documents. This manual is written in Chinese.
| Output | RS-485, MODBUS/RTU |
|---|---|
| Measuring Method | Non contact electromagnetic principle |
| Range | 0 ~ 70PSU |
| Accuracy | ±1%F.S. or ±0.2PSU (Below 10psu) |
| Resolution | 0.1PSU |
| Working Environment | 0 ~ 65 ℃ ; < 0.6MPa |
| Calibration Method | Two point calibration |
| Response Time | 10 seconds |
| Temperature Compensation | Automatic temperature compensation(PT1000) |
| Power Supply | 12-24VDC±10%, 10mA |
| Size | Diameter 30mm; Length 185.5mm; Protection level IP68; Water depth 20 meters |
| Service Life | 3 years or above |
The sensor measures salinity from 0 to 70 PSU.
It uses a built-in PT1000 temperature sensor for automatic compensation.
The default baud rate is 9600 bps.
It is recommended to clean the sensor probe every 30 days.