User manual for CT3510 smart insulation resistance tester covering operation, specs, battery charging, safety and accessories.
This manual provides instructions for operating the CT3510 smart insulation resistance tester, including voltage selection, testing procedures, battery charging, and safety precautions. It details technical specifications such as output voltages (DC500V to DC10000V) and measurement ranges (100KΩ to 10TΩ). The device features automatic absorption ratio and polarization index testing, and includes data logging with time stamps. Safety warnings emphasize avoiding contact with live test leads and ensuring proper discharge before disconnecting cables. This manual is written in Chinese.
| Output Voltage | DC500V, DC1000V, DC2500V, DC5000V, DC10000V |
|---|---|
| Resistance Display Range | 100KΩ to 10TΩ |
| Discharge Time | <250ms/μF from 5000V to 50V |
| Protection Rating | IP65 (cover closed), IP40 (cover open) |
| Operating Temperature | 0℃–40℃ |
| Operating Humidity | ≤70%RH |
| Power Supply | AC 220V±10% or internal lithium battery (DC) |
| Dimensions | 280mm × 210mm × 100mm |
It can test at DC500V, DC1000V, DC2500V, DC5000V, and DC10000V.
Charging takes approximately 3 hours using the included charger.