Chondrex 6705 Mouse Interleukin 1 Beta Detection Kit instructions for quantitative ELISA testing of cell culture media, serum, and plasma samples.
The Chondrex 6705 Mouse Interleukin 1 Beta Detection Kit is a quantitative ELISA kit for measuring mouse IL-1b in cell culture media, serum, and plasma samples. The kit includes standards, detection antibody, dilution buffers, streptavidin peroxidase, TMB and chromogen dilution buffer, stop solution, wash buffer, and a capture antibody coated 96-well ELISA plate.
The document gives notes before using the assay, sample preparation steps, standard dilutions, plate layout, washing and reagent incubation steps, and calculation of results. It specifies a recommended standard range of 16-1000 pg/ml and reading OD values at 450 nm, with a 630 nm filter as a reference.
| Sample types | Cell culture media, serum, and plasma |
|---|---|
| Recommended standard range | 16-1000 pg/ml |
| Plate format | 96-well; 8-well strips x 12 |
| Readout | 450 nm/630 nm |
| Storage temperature | -20°C |
| Inter-assay CV | 7.0%, 7.9%, 5.2% |
| Intra-assay CV | 9.7%, 9.6%, 4.6% |
| Spike test | 91%, 95%, 100% |
The kit is for cell culture media, serum, and plasma samples.
The recommended standard range is 16-1000 pg/ml.
Dilute samples at least 1:1 with Solution B.
Read the OD values at 450 nm.