Short user manual for ChiralTek ChiralCE chromatography columns covering description, operating limits, care, specifications and contact information.
The ChiralTek ChiralCE short user manual provides an overview of the column construction, the unique R1 and R2 substituent groups, and the range of applications such as normal‑phase, reverse‑phase and organic‑polar mobile phases. It also explains the recommended solvent pre‑conditioning steps and flow‑rate limits for both HPLC and UPLC systems.
A detailed specification table lists the particle size, column dimensions, and maximum pressure for each ChiralCE model. The manual includes practical maintenance instructions, storage recommendations, and contact details for the Singapore‑based technical team.
| Particle size | 2–3 µm |
|---|---|
| Maximum pressure | < 800 bar |
| Operating temperature | 0 – 40 °C |
Approximately 0.1 mL/min to avoid exceeding pump pressure limits.
Replace the solvent in the column with methanol for reverse‑phase or isopropanol for normal‑phase storage.
The column can operate up to less than 800 bar (≈11 600 psi).