This user manual covers the installation, operation, maintenance, and calibration of the Campbell Scientific ClariVUE™10 Side-Scatter Turbidity Sensor for…
The ClariVUE™10 is a submersible turbidity sensor that uses side-scatter technology to measure suspended solids and turbidity in water. It is designed for applications ranging from water quality monitoring in aquatic ecosystems to sediment transport and dredge monitoring. The sensor utilizes an near-infrared (NIR) laser and a photodiode to detect light scattered from suspended particles, with the detector positioned at 90 degrees to the emitter.
| Measurement range | 0 to 4000 FNU |
|---|---|
| Measurement response time | 9 to 22 s |
| Resolution | 0.2 FNU |
| Accuracy | ±2% or 0.5 FNU |
| Operating temperature range | –2 to 40°C |
| Power requirements | 9.6 to 18 VDC |
| Maximum depth | 30 m (100 ft) |
| Standards | ISO 7027-1:2016 |
Campbell Scientific recommends factory recalibration every two years or when it varies by more than 5% in standard buffer.
The measurement response time is between 9 and 22 seconds depending on the SDI-12 command used.
Yes, multiple SDI-12 sensors can connect to the same datalogger control or universal terminal, but each must have a unique SDI-12 address.