BioTek Synergy 2 Setup Guide for Predictor hERG FP Assay

Setup guide for configuring the BioTek Synergy 2 Multi-Mode Microplate Reader for the Predictor hERG FP Assay, including optics selection and Gen5 software…


BioTek Synergy 2 Setup Guide for Predictor hERG FP Assay - cover page
Brand
BioTek Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
999 KB
Published
06 November, 2026
Updated
06 November, 2026
MD5 checksum
1AA1C1AD570025F1F26A7DA426C01F25

About this manual

This document provides a setup guide for the BioTek Instruments Synergy™ 2 Multi-Mode Microplate Reader specifically for use with the Predictor™ hERG FP Assay. It details the recommended optics configuration, including excitation and emission filters, and provides step-by-step instructions for configuring the Gen5 software protocol.

The guide also outlines the experimental procedure for preparing the assay plate, including compound dilution, control preparation (negative, positive, free tracer, and blank), and the incubation steps required before reading the plate.

What's inside

  • Recommended Optics
  • Instrument Setup
  • Gen5 Software Configuration
  • Plate Layout
  • Compound Preparation
  • Control Preparation
  • Assay Procedure

Specifications

Excitation Filter530/25
Emission Filter590/35
Dichroic Mirror570 (or 550)
Read Speed150 measurements

Frequently asked questions

How should the read speed be set for best results?

The read speed should be set to 150 measurements; using fewer measurements may decrease read time but could adversely affect assay performance.

What specific assay is this setup guide intended for?

This guide demonstrates the setup of the BioTek Instruments Synergy™ 2 Multi-Mode Microplate Reader for the Predictor™ hERG FP Assay.

What volume of compound should be added per well?

You must add 5 µl of compound per well to the untreated Corning 3677 black polystyrene 384-well plate.

How long should the plate be incubated before reading?

After shaking the plate on an orbital plate shaker for 30 seconds, you must incubate it for 2 hours before reading.