Setup guide for configuring the BioTek Synergy 2 Multi-Mode Microplate Reader for the Predictor hERG FP Assay, including optics selection and Gen5 software…
This document provides a setup guide for the BioTek Instruments Synergy™ 2 Multi-Mode Microplate Reader specifically for use with the Predictor™ hERG FP Assay. It details the recommended optics configuration, including excitation and emission filters, and provides step-by-step instructions for configuring the Gen5 software protocol.
The guide also outlines the experimental procedure for preparing the assay plate, including compound dilution, control preparation (negative, positive, free tracer, and blank), and the incubation steps required before reading the plate.
| Excitation Filter | 530/25 |
|---|---|
| Emission Filter | 590/35 |
| Dichroic Mirror | 570 (or 550) |
| Read Speed | 150 measurements |
The read speed should be set to 150 measurements; using fewer measurements may decrease read time but could adversely affect assay performance.
This guide demonstrates the setup of the BioTek Instruments Synergy™ 2 Multi-Mode Microplate Reader for the Predictor™ hERG FP Assay.
You must add 5 µl of compound per well to the untreated Corning 3677 black polystyrene 384-well plate.
After shaking the plate on an orbital plate shaker for 30 seconds, you must incubate it for 2 hours before reading.