User manual for Automatic Science ASE-12, ASE-16 and ASE-24 solid phase extraction units, covering structure, specifications, setup and operation.
This is the Chinese user manual for the Automatic Science (Tianjin) Instrument Co., Ltd. ASE Series solid phase extraction (SPE) device, covering the ASE-12, ASE-16 and ASE-24 models. It explains the principle of solid phase extraction, the design features of the unit, and lists the main technical parameters such as overall dimensions, glass vacuum chamber size, pressure resistance, sample capacity, pressure range, pressure control accuracy and maximum flow.
The manual then walks through unpacking, the shipping list and acceptance testing with the recommended AP-02B vacuum pump, followed by instructions for adjusting the universal test tube tray to the tube length and diameter and for running an extraction: connecting the pump, inserting SPE columns, setting the negative pressure, loading samples, adjusting flow with the stopcocks and releasing the vacuum. A guide to SPE column and sorbent selection, a warranty statement and a customer reply form are listed at the end of the table of contents. This manual is written in Chinese.
| Overall dimensions (ASE-12, ASE-16) | 280 W x 122 L x 186 H mm |
|---|---|
| Overall dimensions (ASE-24) | 287 W x 125 L x 188 H mm |
| Glass vacuum chamber dimensions (ASE-12, ASE-16) | 240 W x 85 L x 140 H mm |
| Glass vacuum chamber dimensions (ASE-24) | 247 W x 88 L x 142 H mm |
| Pressure resistance | Better than 80 kPa (65 kPa recommended) |
| Samples processed simultaneously | ASE-12: 12; ASE-16: 16; ASE-24: 24 |
| Stable pressure range | 0 to 80 kPa, adjustable |
| Pressure control accuracy | ±5 kPa |
| Maximum flow | 10 L/min |
| Test tube rack sizes | Φ11, Φ13, Φ15 mm; length 105 mm and below |
| Recommended pump | AP-02B |
It is a sample preparation technique that uses a solid sorbent to adsorb target compounds from a liquid sample, separating them from the sample matrix and interfering compounds, after which they are eluted with a solvent or by thermal desorption.
The manual recommends the AP-02B adjustable regulated pump, which has an LED pressure display and a built-in pressure sensor.
The pressure should preferably not exceed -65 kPa (or -20 Hg) so that the glass vacuum chamber is not damaged.
Set the tray height so the distance from its surface to the top of the glass chamber is about 10 mm greater than the tube length, then lock it with three Ω-shaped clips in the nearest locating slots.