ARK Linezolid Assay Product Leaflet

ARK Linezolid Assay is a homogeneous enzyme immunoassay for quantifying linezolid in human serum on automated clinical chemistry analyzers.


ARK Linezolid Assay Product Leaflet - cover page
Brand
ARK
Category
medical technology
Document type
Product Leaflet
Language
English
Pages
4
File format
PDF
File size
783 KB
Published
03 November, 2026
Updated
03 November, 2026
MD5 checksum
176EE11702C0461E07FBC4280495DEA3

About this manual

The ARK Linezolid Assay is a homogeneous enzyme immunoassay designed for quantitative determination of linezolid in human serum. It uses competition between drug in the specimen and enzyme-labeled linezolid for antibody binding. Results are measured spectrophotometrically as NAD to NADH conversion rate. The assay is intended for therapeutic drug monitoring to help ensure appropriate therapy. It is for in vitro diagnostic use only and requires prescription.

The assay includes reagents, calibrators, and controls. Specimens must be serum, and trough (pre-dose) samples are preferred. Storage and handling instructions are provided, including temperature limits and precautions for sodium azide. Results are reported in μg/mL or μmol/L, with conversion factors given. The assay has a measurement range of 0.75–30.00 μg/mL and can be diluted for higher concentrations. Performance characteristics are provided for the Beckman Coulter AU680 System, including limits of blank, detection, and quantitation.

What's inside

  • Intended Use
  • Summary and Explanation of the Test
  • Principles of the Procedure
  • Reagents
  • Warnings and Precautions
  • Specimen Collection and Preparation
  • Procedure
  • Results and Expected Values
  • Limitations of the Procedure
  • Expected Values

Specifications

Measurement Range0.75 – 30.00 μg/mL
Limit of Blank (LoB)0.003 μg/mL
Limit of Detection (LoD)0.071 μg/mL
Limit of Quantitation (LoQ)0.75 μg/mL
Conversion FactorMultiply μg/mL by 2.964 to convert to μmol/L

Frequently asked questions

What is the measurement range of the ARK Linezolid Assay?

The measurement range is 0.75 – 30.00 μg/mL.

What are the units for reporting results?

Results are reported in μg/mL or μmol/L. To convert μg/mL to μmol/L, multiply by 2.964.

What is the limit of quantitation (LoQ) for the assay?

The LoQ is 0.75 μg/mL.