ARK Linezolid Assay is a homogeneous enzyme immunoassay for quantifying linezolid in human serum on automated clinical chemistry analyzers.
The ARK Linezolid Assay is a homogeneous enzyme immunoassay designed for quantitative determination of linezolid in human serum. It uses competition between drug in the specimen and enzyme-labeled linezolid for antibody binding. Results are measured spectrophotometrically as NAD to NADH conversion rate. The assay is intended for therapeutic drug monitoring to help ensure appropriate therapy. It is for in vitro diagnostic use only and requires prescription.
The assay includes reagents, calibrators, and controls. Specimens must be serum, and trough (pre-dose) samples are preferred. Storage and handling instructions are provided, including temperature limits and precautions for sodium azide. Results are reported in μg/mL or μmol/L, with conversion factors given. The assay has a measurement range of 0.75–30.00 μg/mL and can be diluted for higher concentrations. Performance characteristics are provided for the Beckman Coulter AU680 System, including limits of blank, detection, and quantitation.
| Measurement Range | 0.75 – 30.00 μg/mL |
|---|---|
| Limit of Blank (LoB) | 0.003 μg/mL |
| Limit of Detection (LoD) | 0.071 μg/mL |
| Limit of Quantitation (LoQ) | 0.75 μg/mL |
| Conversion Factor | Multiply μg/mL by 2.964 to convert to μmol/L |
The measurement range is 0.75 – 30.00 μg/mL.
Results are reported in μg/mL or μmol/L. To convert μg/mL to μmol/L, multiply by 2.964.
The LoQ is 0.75 μg/mL.