This manual provides instructions for the Applygen E1016 kit, an enzymatic assay for measuring free cholesterol in tissue and cells.
The Applygen E1016 Tissue Free Cholesterol Assay Kit is designed for the determination of free cholesterol in solid tissues and cells using an enzymatic method. This approach avoids the use of toxic organic solvents and complex steps, offering a simple and sensitive procedure suitable for animal tissues and cultured cells. The kit utilizes a WHO-recommended enzymatic method combined with the GPO Trinder reaction to measure cholesterol concentration. This manual is written in Chinese.
| Storage Temperature | 4 ºC |
|---|---|
| Best Wavelength | 550nm |
| Alternative Wavelengths | 570nm, 530nm, 490nm |
| Reaction Temperature | 37º C or 25º C |
| Reaction Time | 20 minutes |
| Kit Capacity (Microplate) | 105 tests |
| Kit Capacity (Cuvette) | 30 tests |
The reaction should be performed at 37º C or 25º C room temperature for 20 minutes.
The best working wavelength is 550nm, though 570nm, 530nm, or 490nm can be used if 550nm is unavailable.
Interference may occur with Vitamin C >0.18g/L, Hemoglobin >2g/L, DTT, mercaptoethanol, and high concentration EDTA.