AEC Q200-002 Rev B Human Body Model ESD Test Specification

This AEC document specifies the procedure for determining Electrostatic Discharge (ESD) sensitivity of passive components using the Human Body Model.


AEC Q200-002 Rev B Human Body Model ESD Test Specification - cover page
Brand
AEC
Document type
Datasheet
Language
English
Pages
14
File format
PDF
File size
127 KB
Published
16 October, 2026
Updated
16 October, 2026
MD5 checksum
BC457FCFCEF0F86213D92D60FC7B12BA

About this manual

This specification establishes a reliable and repeatable procedure for testing the Electrostatic Discharge (ESD) sensitivity of passive electronic components. It defines the requirements for component failure, device under test (DUT), and the electrostatic discharge simulator used in the evaluation.

The document details the necessary equipment, including ESD pulse simulators, oscilloscopes with specific bandwidth requirements, coaxial targets, and ground planes. It provides comprehensive instructions for qualifying and verifying the ESD simulator through direct contact and air discharge methods to ensure waveform compliance before testing.

What's inside

  • Scope
  • Terms and Definitions
  • Equipment Requirements
  • ESD Simulator Qualification
  • Simulator Charge Verification
  • Waveform Parameter Requirements
  • Direct Contact Discharge
  • Air Discharge

Specifications

Oscilloscope Minimum Bandwidth (Analog)1.0GHz
Oscilloscope Sampling Rate (Digital)2 gigasamples per second
Ground Plane Minimum Thickness1mm
Ground Plane Minimum Area1m2
Ground Strap Maximum Length1m
Ground Strap Maximum Inductance2µH
Electrometer Probe Input Impedance≥ 100 Gigohm
Simulator Calibration Interval6 months

Frequently asked questions

What is the purpose of this specification?

The purpose is to establish a reliable and repeatable procedure for determining PASSIVE COMPONENT HBM ESD sensitivity.

How often must the ESD simulator be qualified?

The simulator must be fully qualified during initial acceptance testing, whenever serviced, and at least every six months.

Download PDF (127 KB)