User manual for Adam Equipment's ABI-1 ionizer, covering operation, safety, maintenance, and applications for neutralizing static charges in weighing…
This user manual provides instructions for operating the Adam Equipment ABI-1 ionizer, designed to neutralize static charges on materials during weighing. It includes safety regulations, technical specifications, maintenance procedures, and warranty information. The device uses high-voltage needles to generate positive and negative ions that neutralize static charges, improving weighing accuracy. It is intended for use with electronic scales only and must be used indoors with proper ventilation due to ozone production. This manual is written in Spanish.
| Ionization Distance | approx. 5 - 40 cm |
|---|---|
| Discharge Time | 9s / 5cm, 13s / 10cm, 100s / 40cm (with fan on) |
| Ozone Concentration | 0 ~ 0.05 ppm (2cm from ion source) |
| Air Volume | 0.06cm³/min |
| Operating Temperature | 0 – 50°C |
| Input Voltage | AC 100-240V, 50/60Hz |
| Secondary Power | DC 12V, 500mA |
| Dimensions | 11 x 10.5 x 4.5 cm |
| Weight | 310g |
It neutralizes static charges on materials during weighing by generating positive and negative ions.
Do not use in explosive environments, avoid high humidity, do not disassemble, use only the provided power adapter, and ensure proper ventilation due to ozone production.
Clean the ion source after 100 hours of use and replace it after 10,000 hours.