This document describes the features and characteristics of the high pressure digestion tank produced by Shanghai XinNuo Instrument and Equipment Co., Ltd.
This is the Chinese user manual for the Xinnuo (Shanghai Xinnuo Instrument Equipment Co., Ltd.) GCXJ Series high pressure digestion tank, a steel-lined PTFE (F4) sealed digestion vessel also described in the text as a high temperature digestion tank, high pressure reaction vessel, PTFE digestion tank or hydrothermal synthesis vessel. The document explains that the vessel consists of a 1Cr18Ni9Ti stainless steel outer sleeve and a PTFE inner cup, and lists typical uses in sample preparation for food, biology, geology, metallurgy, environmental protection, commodity inspection, chemical, military and nuclear laboratories, including heavy metal determination and organic or hydrothermal synthesis. It also names the national standards the product is said to follow, such as GB/T5009, GB/T14962, GB/T6609, GB/T11914 and SN/T2004.1-2005 / SN/T2004.2-2005.
The manual gives technical characteristics (materials, operating temperature, pressure range and available volumes), a section on the properties of the PTFE liner, the working principle, and step-by-step instructions for use before, during and after digestion. The instructions describe soaking the PTFE cup in acid, loading sample and solvent, assembling the cup in the steel sleeve, heating in an oven at 100 °C for one hour and then at the required temperature for two hours, cooling before opening, and cleaning and dry storage. Warnings advise against microwave oven use, against temperatures of 180 °C or above, and against use in a muffle furnace, and note that the solvent volume is usually one third to one fifth of the inner cup. This manual is written in Chinese.
| Material | PTFE (F4) inner liner with 1Cr18Ni9Ti stainless steel outer sleeve |
|---|---|
| Operating temperature | Use in an oven up to 200 °C; working temperature recommended not to exceed 180 °C |
| Pressure range | ≤ 3 MPa (about 30 kg/cm²) |
| Available volumes | 4, 10, 15, 20, 25, 30, 50, 60, 90, 100, 200 and 500 ml |
| PTFE service temperature | -200 to +250 °C |
| PTFE friction coefficient | 0.04 |
| Optional PFA inner cup | Sizes of 30 ml and below can be made with a PFA inner cup |
Heat the assembled tank in an oven: hold at 100 °C for one hour, then raise to the required temperature for two hours. Most samples digest completely between 80 °C and 140 °C, and temperatures of 180 °C or above are not recommended.
No. The manual states it is forbidden to place the tank in a microwave oven; use in a muffle furnace is also not recommended.
The amount depends on the experiment, but is generally one third to one fifth of the inner cup.
Listed sizes are 4, 10, 15, 20, 25, 30, 50, 60, 90, 100, 200 and 500 ml, and custom sizes and PTFE inner cups can be made from a sample or drawing.