PV411 multifunctional hand pump can generate vacuum, air pressure and hydraulic pressure.
The PV411 is a multifunction hand pump that generates vacuum, pneumatic pressure and hydraulic pressure. The pneumatic side has a working-mode switching valve that changes between pressure and vacuum, a pressure handle for pumping and a volume adjuster for fine pressure adjustment. The hydraulic side uses the volume adjuster to build pressure, with a reservoir on top of the pump supplying the hydraulic medium (water or oil) and an over-pressure protection valve that can be set between 3 and 70 MPa.
The manual gives on-site operating instructions, safety notes, a list of abbreviations and symbols, connection details for the pressure ports, filling and venting of the hydraulic system, setting of the over-pressure protection valve, changing and cleaning of the pressure medium, and a troubleshooting section. Stated specifications include a pneumatic range of 0 to 6 MPa, a vacuum range of 0 to -0.095 MPa, a hydraulic range of 0 to 70 MPa, a 1/4 BSP female pressure port, a weight of 1.0 kg and dimensions of 260×135×95 mm; the pump head can be rotated 140°. This manual is written in Chinese.
| Pneumatic pressure range | 0 to 6 MPa |
|---|---|
| Vacuum range | 0 to -0.095 MPa (at 100 kPa, varies with local atmospheric pressure) |
| Hydraulic pressure range | 0 to 70 MPa |
| Maximum working pressure | 70 MPa |
| Hydraulic medium | Distilled water or mineral-based oil |
| Pressure port | 1/4 BSP female thread |
| Weight | 1.0 kg |
| Dimensions | 260 × 135 × 95 mm |
| Over-pressure protection valve setting range | 3 to 70 MPa |
The pneumatic range is 0 to 6 MPa, the vacuum range is 0 to -0.095 MPa and the hydraulic range is 0 to 70 MPa, with a maximum working pressure of 70 MPa.
Turn the pressure/vacuum switching valve fully clockwise for pressure; turn it fully counter-clockwise for vacuum, and leave it in the middle position to vent.
Build an initial pressure higher than the required set point, loosen the valve's lock screw and turn it clockwise, slowly release the pressure to the required value, then tighten the lock screw.
Use distilled water or mineral-based oil, and do not mix pressure media.