Malvern ZEN1010 High Concentration Cell User Manual

This document gives a brief overview of the Malvern Zetasizer Nano High concentration cell. It primarily describes how to use, insert and clean the cell to ensure reliable and consistent measurements.


Malvern ZEN1010 High Concentration Cell User Manual - cover page
Brand
Malvern
Category
Analyzer
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
546 KB
Published
29 November, 2023
Updated
29 November, 2023
MD5 checksum
EC251D63E37E7A825F493E6F5D40305F

About this manual

The Malvern ZEN1010 High Concentration Cell is intended primarily for measuring zeta potential of concentrated aqueous samples with the Zetasizer Nano. It can be used with the MPT-2 for automated titrations and consists of a high precision optical measurement block held within electrode chambers in an outer cuvette-sized casing.

The document explains selecting the cell in the relevant Cell SOP window so the Zetasizer software configures settings and parameters. It also covers slow sample filling, checking for air bubbles, inserting the cell with the metal face toward the front of the instrument, and general or intensive cleaning.

What's inside

  • Description
  • In the SOP
  • Cleaning the High concentration cell
  • General cleaning
  • Filling the cell
  • Inserting the Cell
  • Intensive cleaning
  • Chemical compatibility

Specifications

Intended measurementZeta potential of concentrated aqueous samples
Automated titration useMPT-2 for automated titrations
Cell constructionHigh precision optical measurement block held within electrode chambers
CasingOuter cuvette-sized casing assembly
Supplied tubing1/32" internal bore silicon tubing with Luer fittings
Measurement block materialQuartz
Casing materialsDelrin / Stainless steel 316

Frequently asked questions

What is the Malvern ZEN1010 High Concentration Cell used for?

It is intended primarily for measuring zeta potential of concentrated aqueous samples.

How do I fill the High Concentration Cell?

Inject the sample slowly until the liquid reaches the bottom of the luer outlet, check that no air bubbles form, and insert a cell stopper in each port.

How should the cell be inserted into the instrument?

The metal face of the cell must face the front of the instrument to ensure good thermal contact between cell and instrument.

How should the High Concentration Cell be cleaned before measurement?

Rinse it by flushing through with copious amounts of deionized water; external surfaces of the assembled cell can be wiped with a weak soap solution.

Download PDF (546 KB)