This document is the instruction manual for the VACTEST TPP900 vacuum measurement equipment mobile pressure gauge.
The VACTEST TPP900 is a mobile vacuum gauge that measures total pressure from 1200 to 5 x 10^-4 mbar using combined piezoresistive and Pirani sensors. It can be operated under vacuum inside a vacuum chamber and stores up to 2000 measurements.
This instruction manual provides safety information, product description, transport and storage, installation conditions, vacuum and electrical connections, USB interface, operation, data recording, PC communication, maintenance, troubleshooting, accessories, and technical data.
| Measurement principle | Piezoresistive / Pirani |
|---|---|
| Measuring range mbar | 1200 … 5 x 10-4 |
| Overpressure limit bar abs. | 2 |
| Measurement uncertainty | ±0.3 % (1200 … 40 mbar) |
| Resolution | 1 mbar (1200 … 1000 mbar) |
| Reaction time s | <1 |
| Serial Interface | USB |
| Supply voltage | 9 V block battery or 15 VDC external |
| Max. battery operating time h | 100 |
| Protection Class | IP 40 |
It measures total pressure from 1200 to 5 x 10^-4 mbar.
It can store up to 2000 measurements in the vacuum gauge.
9 Volt AlMn block battery type 6LR 61 or 9 Volt Lithium block battery.
The protection class is IP 40.