InstruTech CCM501 Hornet Cold Cathode Ionization Vacuum Gauge User Manual

InstruTech® Cold Cathode Ionization Vacuum Gauge CCM501 Module The Hornet™ User Manual


InstruTech CCM501 Hornet Cold Cathode Ionization Vacuum Gauge User Manual - cover page
Brand
Instrutech
Category
Monitoring Equipment
Document type
User Manual
Language
English
Pages
38
File format
PDF
File size
1.7 MB
Published
24 November, 2023
Updated
24 November, 2023
MD5 checksum
21C28C9C5C6ED2D1C0B2A9CD5CE8C156

About this manual

The InstruTech CCM501 Hornet is a module designed to provide signal conditioning for a cold cathode vacuum gauge, transforming it into a complete measuring instrument. It features a log-linear analog output, RS485 serial communications, a setpoint relay, and a built-in OLED graphical display for user interface and setup.

The device measures vacuum pressure by ionizing gas molecules and measuring the resulting ion current, with a measurement range from 1 x 10-9 to 1 x 10-2 Torr. The design includes an internal O-ring to facilitate easy disassembly for cleaning.

What's inside

  • Introduction / General Information
  • Important Safety Information
  • Installation
  • Bakeout
  • Setup and Operation
  • Using the gauge with different gases
  • Analog Output
  • RS485 serial communications
  • Service

Specifications

Measurement range1 x 10-9 to 1 x 10-2 Torr
Accuracy (N2 typical)± 30% of reading (1 x 10-8 to 1 x 10-2 Torr)
Repeatability (typical)± 5% of reading
Input power20 to 28 Vdc, 7.2 W
Operating temperature0 oC to + 40 oC
Bakeout temperature150 oC (sensor only)

Frequently asked questions

What is the measurement range of the CCM501 gauge?

The measurement range is 1 x 10-9 to 1 x 10-2 Torr, 1.3 x 10-9 to 1.3 x 10-2 mbar, or 1.3 x 10-7 Pa to 1.3 Pa.

What materials are exposed to gases in the sensor?

The materials exposed to gases are 304 stainless steel, ceramic, and Viton® O-rings.