Standard Operating Procedure for Operation of Analytical Balances

This document describes the procedure for accurate weighing (directly or indirectly) of samples using a digital analytical balance.


Standard Operating Procedure for Operation of Analytical Balances - cover page
Document type
Other
Language
English
Pages
2
File format
PDF
File size
143 KB
Published
24 November, 2023
Updated
24 November, 2023
MD5 checksum
4BE5267AD30FA4885F59A22DBA97365C

About this manual

This Standard Operating Procedure (SOP-BA07), effective 23/03/2021, outlines how to operate digital analytical balances in the TRACES Centre and undergraduate laboratories. It applies to devices including Sartorius, Ohaus, and Mettler-Toledo balances, with the user responsible and the TRACES Manager or Course Instructor accountable.

The procedure covers leveling and zeroing the balance, direct weighing, indirect weighing by difference, and cleaning up after use. Users should consult TRACES staff, teaching assistants, or the lab manual for the appropriate weighing option and clarifications.

What's inside

  • Introduction
  • Referenced Documents
  • Equipment
  • Procedures
  • Balance Operation
  • Before you start
  • Direct Weighing
  • Indirect Weighing (Weighing by difference)
  • Cleaning up

Frequently asked questions

What is the purpose of this procedure?

It outlines how to operate digital analytical balances in the TRACES Centre and undergraduate laboratories.

How do you zero the analytical balance?

Close all the balance doors, press the button labelled ‘0’, and wait until the unit shows all zeros.

What is indirect weighing?

It determines sample mass by weighing a tared vessel before and after removing a determined amount of sample.

What should you do after weighing is completed?

Clean up any spilled chemicals and use the provided brush to remove particulates from the pan area.