This document describes the procedure for accurate weighing (directly or indirectly) of samples using a digital analytical balance.
This Standard Operating Procedure (SOP-BA07), effective 23/03/2021, outlines how to operate digital analytical balances in the TRACES Centre and undergraduate laboratories. It applies to devices including Sartorius, Ohaus, and Mettler-Toledo balances, with the user responsible and the TRACES Manager or Course Instructor accountable.
The procedure covers leveling and zeroing the balance, direct weighing, indirect weighing by difference, and cleaning up after use. Users should consult TRACES staff, teaching assistants, or the lab manual for the appropriate weighing option and clarifications.
It outlines how to operate digital analytical balances in the TRACES Centre and undergraduate laboratories.
Close all the balance doors, press the button labelled ‘0’, and wait until the unit shows all zeros.
It determines sample mass by weighing a tared vessel before and after removing a determined amount of sample.
Clean up any spilled chemicals and use the provided brush to remove particulates from the pan area.