The PHG-206 Online PH Sensor produced by Yantai Chemins Instrument Co., Ltd. is a high-precision and high-stability sensor. It adopts imported chips and sensors, with high precision, good stability, fast response and strong anti-interference ability. It is suitable for various industrial occasions.
The PHG-206 is an online pH sensor from Chemins for environmental water quality monitoring, acid/alkali/salt solutions, chemical reaction processes and industrial production processes. It provides RS-485 (Modbus/RTU) output and can connect to devices such as PLCs, DCS systems, industrial control computers, general-purpose controllers, paperless recording instruments or touch screens.
The manual describes the sensor's technical parameters and dimensions, installation and electrical connection, use and maintenance, calibration, quality and service information, accessories, and Modbus data communication registers.
| Model | PHG-206 |
|---|---|
| Measuring range | 0-14 pH |
| Resolution | 0.01 pH |
| Accuracy | ±0.1 pH |
| Working temperature | 0-65 C |
| Working pressure | <0.2 MPa |
| Temperature compensation | Auto temperature compensation (NTC) |
| Power supply | 12-24 VDC ±10% |
| Signal output | RS-485 (Modbus/RTU) |
| Wetted material | PPR |
| Mounting method | Immersion mounting, 3/4 NPT thread |
| Cable length | 5 meters, other lengths can be customized |
| Calibration | Two-point calibration |
| Power consumption | <0.5 W |
| Protection grade | IP68 |
| Cable outer diameter | Φ6 mm |
Red is the power cord, black is ground, blue is 485A, white is 485B, and the bare wire is the shielded wire.
The default format is 9600, n, 8, 1.
The default address is 6, with a data range of 1-127.
Place the sensor in a pH 6.86 standard solution, wait 3 to 5 minutes, and perform zero calibration if the displayed value is not 6.86.