Chemins PHG-206 Online pH Sensor User Manual

The PHG-206 Online PH Sensor produced by Yantai Chemins Instrument Co., Ltd. is a high-precision and high-stability sensor. It adopts imported chips and sensors, with high precision, good stability, fast response and strong anti-interference ability. It is suitable for various industrial occasions.


Chemins PHG-206 Online pH Sensor User Manual - cover page
Brand
Chemins
Category
Other products
Document type
User Manual
Language
English
Pages
11
File format
PDF
File size
254 KB
Published
22 November, 2023
Updated
22 November, 2023
MD5 checksum
7E46F8F5B4FD4443BB91066485CEBB42

About this manual

The PHG-206 is an online pH sensor from Chemins for environmental water quality monitoring, acid/alkali/salt solutions, chemical reaction processes and industrial production processes. It provides RS-485 (Modbus/RTU) output and can connect to devices such as PLCs, DCS systems, industrial control computers, general-purpose controllers, paperless recording instruments or touch screens.

The manual describes the sensor's technical parameters and dimensions, installation and electrical connection, use and maintenance, calibration, quality and service information, accessories, and Modbus data communication registers.

What's inside

  • Application environment
  • Technical parameters
  • Dimensional drawing
  • Installation
  • Electrical connection
  • Use and maintenance
  • Calibration
  • Quality assurance
  • Accessories and spare parts
  • Data communication

Specifications

ModelPHG-206
Measuring range0-14 pH
Resolution0.01 pH
Accuracy±0.1 pH
Working temperature0-65 C
Working pressure<0.2 MPa
Temperature compensationAuto temperature compensation (NTC)
Power supply12-24 VDC ±10%
Signal outputRS-485 (Modbus/RTU)
Wetted materialPPR
Mounting methodImmersion mounting, 3/4 NPT thread
Cable length5 meters, other lengths can be customized
CalibrationTwo-point calibration
Power consumption<0.5 W
Protection gradeIP68
Cable outer diameterΦ6 mm

Frequently asked questions

What wire colors are used for the PHG-206 electrical connection?

Red is the power cord, black is ground, blue is 485A, white is 485B, and the bare wire is the shielded wire.

What is the default Modbus communication format?

The default format is 9600, n, 8, 1.

What is the default device address?

The default address is 6, with a data range of 1-127.

How is zero calibration performed?

Place the sensor in a pH 6.86 standard solution, wait 3 to 5 minutes, and perform zero calibration if the displayed value is not 6.86.