752S/752Pro/752Bas UV-VIS Spectrophotometer is a simple and easy-to-use spectrophotometer that can perform transmittance, absorbance and concentration direct reading measurements in the wavelength range of 190nm to 1100nm. It is widely used in medical and health, clinical testing, biochemistry, petrochemical, environmental monitoring, quality control and other departments for qualitative and quantitative analysis.
The Shanghai Lengguang 752S/752Pro/752 Bas is a UV-VIS spectrophotometer for transmittance, absorbance, and concentration measurements from 190 nm to 1100 nm. It has a 4-digit LED display, C-T grating monochromator, 4-position sample holder, automatic zero and 100%T adjustment, concentration factor and direct concentration readout, and an RS-232C serial interface.
The manual covers installation requirements, basic operation such as wavelength adjustment and scale selection, application procedures including transmittance and absorbance measurements, and maintenance tasks such as lamp replacement and common fault checks. The three models differ in transmittance accuracy, repeatability, spectral bandwidth, and stray light. This manual is written in Chinese.
| Wavelength range | 190 nm to 1100 nm |
|---|---|
| Optical system | Diffraction grating C-T monochromator |
| Light source | Deuterium and tungsten lamps |
| Receiver | Photocell |
| Wavelength accuracy | ±2.0 nm |
| Wavelength repeatability | 0.5 nm |
| Transmittance accuracy | ±0.5%T (752Bas/752Pro), ±0.3%T (752S) |
| Transmittance repeatability | 0.2%T (752Bas/752Pro), 0.1%T (752S) |
| Spectral bandwidth | 4±0.8 nm (752Bas/752Pro), 2±0.4 nm (752S) |
| Stray light | 0.2%T at 220 nm (752Bas/752Pro), 0.1%T at 220 nm (752S) |
| Stability | Photocurrent drift ≤0.5%T |
| Display scales | T 0.0-199.9%, A -0.3-2.999, F 1-9999, C 0-9999 |
| Power | 220 V ±22 V, 50 Hz ±1 Hz |
| Dimensions | 370 x 320 x 240 mm |
| Weight | Net 7.5 kg; gross 9.5 kg |
It measures from 190 nm to 1100 nm.
It should warm up for 30 minutes after power-on.
220 V ±22 V at 50 Hz ±1 Hz.
Close the light path with an opaque material or close the sample cover, then press the 0% key.