Pre-clinical Testing Guidance for Extracorporeal Shock Wave Lithotripter 107.7

Extracorporeal shock wave lithotripter is a device that uses focused shock waves to break up kidney stones. The main components of the device are: shock wave generator, high pressure generator, control panel, imaging/positioning system, and patient table.


Pre-clinical Testing Guidance for Extracorporeal Shock Wave Lithotripter 107.7 - cover page
Category
Other
Document type
Other
Language
ZH-TW
Pages
7
File format
PDF
File size
1.1 MB
Published
19 November, 2023
Updated
19 November, 2023
MD5 checksum
A93894F556F681D0B0A29CC160330569

About this manual

This document outlines the pre-clinical testing guidelines for Extracorporeal Shock Wave Lithotripters (ESWL), designated as item H.5990. It serves as a reference for medical device manufacturers during product verification registration, covering necessary data and testing items without including clinical trial requirements. The guidelines emphasize that manufacturers must provide complete verification and evaluation data based on their specific product structure, materials, and claimed performance.

The document details the scope of application, which includes focused shock wave lithotripsy equipment and associated patient support and positioning systems, while excluding non-focused devices or accessories like computer planning systems and monitoring equipment. It specifies required product descriptions, specifications (such as power supply, dimensions, and shock wave parameters), and mandatory safety and performance tests, including electrical safety, electromagnetic compatibility, biocompatibility, software validation, and functional tests. This manual is written in Chinese.

What's inside

  • Scope of Application
  • Product Description and Specification
  • Electrical Safety Test
  • Electromagnetic Compatibility Test
  • Biocompatibility Test
  • Software Validation
  • Performance Test
  • References

Specifications

Classification NumberH.5990
Electrical Safety StandardIEC 60601-1 (2012) and IEC 60601-2-36 (2014)
Electromagnetic Compatibility StandardIEC 60601-1-2 (2014) and IEC 60601-2-36 (2014)
Biocompatibility StandardsISO 10993-1 (2009), ISO 10993-5 (2009), ISO 10993-10 (2010)

Frequently asked questions

Does this guidance include clinical trial requirements?

No, this guidance does not include requirements for clinical trials or other data.

What is the Extracorporeal Shock Wave Lithotripter 107.7 used for?

It is a non-invasive device used to fragment kidney or ureteral stones by directing focused shock waves into the body.

What laser hazard classes are covered by the Laser Surgical Instrument guidance?

The guidance applies to laser instruments that have a laser hazard class of Class 3B or Class 4.

Download PDF (1.1 MB)