Extracorporeal shock wave lithotripter is a device that uses focused shock waves to break up kidney stones. The main components of the device are: shock wave generator, high pressure generator, control panel, imaging/positioning system, and patient table.
This document outlines the pre-clinical testing guidelines for Extracorporeal Shock Wave Lithotripters (ESWL), designated as item H.5990. It serves as a reference for medical device manufacturers during product verification registration, covering necessary data and testing items without including clinical trial requirements. The guidelines emphasize that manufacturers must provide complete verification and evaluation data based on their specific product structure, materials, and claimed performance.
The document details the scope of application, which includes focused shock wave lithotripsy equipment and associated patient support and positioning systems, while excluding non-focused devices or accessories like computer planning systems and monitoring equipment. It specifies required product descriptions, specifications (such as power supply, dimensions, and shock wave parameters), and mandatory safety and performance tests, including electrical safety, electromagnetic compatibility, biocompatibility, software validation, and functional tests. This manual is written in Chinese.
| Classification Number | H.5990 |
|---|---|
| Electrical Safety Standard | IEC 60601-1 (2012) and IEC 60601-2-36 (2014) |
| Electromagnetic Compatibility Standard | IEC 60601-1-2 (2014) and IEC 60601-2-36 (2014) |
| Biocompatibility Standards | ISO 10993-1 (2009), ISO 10993-5 (2009), ISO 10993-10 (2010) |
No, this guidance does not include requirements for clinical trials or other data.
It is a non-invasive device used to fragment kidney or ureteral stones by directing focused shock waves into the body.
The guidance applies to laser instruments that have a laser hazard class of Class 3B or Class 4.