This document introduces the basic knowledge of HPLC, including what is HPLC, the working principle of HPLC, the main components of HPLC, the operation method of HPLC, etc.
This document serves as a getting started guide for beginners using the Merii Kato HPLC system. It explains the basic principles of High Performance Liquid Chromatography (HPLC) and details the capabilities of the system, such as identifying unknown substances, simultaneous quantification of multi-components, and separation/purification of mixtures. The guide also covers the preparation of mobile phases and samples, emphasizing filtration and degassing steps to ensure accurate results. This manual is written in Japanese.
HPLC allows for the identification of unknown substances based on known peaks, simultaneous quantification of multiple components, separation and purification of mixtures, and measurement of physicochemical constants.
Check that PA1 and PA2 lights are not red, press mode twice, enter 0, select dual 1, set the flow rate with F-set, check Display 3, turn the heater ON, and turn Pump A ON.
Use a syringe with a capacity of 30ul or more for 20ul samples, ensure no air enters the syringe, insert the needle fully with the plunger down, inject quickly, and then raise the plunger.