WPI's Liquid Ion Exchangers (LIX) are used with the FD223 Electrometer to allow intracellular measurements to be made for cations (hydrogen, potassium and calcium) using the procedures described in this manual.
This instruction manual covers the World Precision Instruments IE Series liquid ion exchangers (IE010, IE190, and IE200) for use with the FD223 Electrometer. It provides cation selectivity coefficients, useful pH ranges, slopes, linear ranges, and references for hydrogen, potassium, and calcium measurements.
The manual also describes how to fabricate ion-selective microelectrodes, including pulling the microelectrode, silanizing the tip, filling the tip with ion exchanger, and checking the finished electrode under a microscope. It includes a 30-day warranty, claims and returns information, and repair instructions.
| Cations | H+, K+, Ca++ |
|---|---|
| IE010 useful pH range | 2-10 |
| IE190 useful pH range | 4-10 |
| IE200 useful pH range | 4-10 |
| IE010 slope | 56 mV |
| IE190 slope | 58 mV |
| IE200 slope | 28 mV |
| IE010 linear range | pH 4-12 |
| IE190 linear range | pK 0-3 |
| IE200 linear range | pCa 1-7 |
| Microelectrode tip diameter | One micron or less |
| Recommended silanizing solution | 5% dimethyldi-chlorosilane in xylene |
| Silanizing exposure time | 1-2 minutes |
| Silanizing bake time | One hour |
| Silanizing bake temperature | 100 °C |
| Ion exchanger column length | 200-500 micron |
| Reference electrolyte | 0.5 M KCl or NaCl |
| Warranty period | 30 days from date of receipt |
| Restocking charge | 25% |
The manual covers the WPI IE010, IE190, and IE200 liquid ion exchangers, which are used with the FD223 Electrometer for intracellular cation measurements.
The IE Series supports intracellular measurements for hydrogen, potassium, and calcium cations.
The general procedure is to pull the microelectrode from a suitable glass capillary, apply a hydrophobic coating to the interior of the electrode tip, and fill the tip with the ion exchanger.
The tip should be checked under a microscope to verify complete filling and confirm that no air bubbles are present in the exchanger column.
WPI warrants the equipment and its components and parts against defects in material and workmanship for 30 days from the date of receipt, subject to the stated conditions.