The GSR D1 Gamma Irradiation Device is a device used for irradiation sterilization. It has the following characteristics: 1. Wide range of irradiation doses to meet the irradiation requirements of different products; 2. Short irradiation time to improve production efficiency; 3. Good irradiation uniformity to ensure consistent product irradiation effect; 4. Simple operation and easy maintenance.
This document provides assembly and installation guidelines for the GSR D1 Gamma Irradiation Device. It specifies environmental requirements for the installation room, including temperature, humidity, and ventilation, as well as structural requirements like floor loading and ceiling height.
| Temperature | 15 - 35 °C |
|---|---|
| Relative air moisture | 45 - 75 % |
| Atmosphere pressure | 860 - 1.060 hPa |
| Air ventilation | 5 renewals of air per hour |
| Height under the ceiling | 2,80 m |
| Entrance width | min. 90 cm |
| Electrical socket | 1 pcs., fused with 16 A |
| Heaviest single part weight | 2,2 tons |
| Assembled weight outside room | 4,3 tons |
The installation room temperature must be between 15 and 35 °C.
Only technicians authorized by Gamma-Service Medical GmbH can carry out the installation.
All doors on the route of transport must have a minimum width of 90 cm.