This document specifies the classification, technical requirements, test methods, inspection rules, marking and instruction manuals, packaging, transportation, storage, etc. of soil moisture (墒情) monitoring instrument products.
This standard specifies the classification, technical requirements, test methods, inspection rules, marking and instruction manual, packaging, transport and storage for soil moisture (soil moisture condition) monitoring instruments. It applies to instruments used directly for soil moisture monitoring in water resources utilization, drought relief, water-cycle research and agricultural irrigation.
The requirements cover general requirements, appearance, operating environmental conditions, basic parameters, electrical performance, mechanical environmental adaptability and reliability. Test methods, inspection rules and related provisions are also included. This manual is written in Chinese.
| Measurement range | 0~50% soil volumetric water content |
|---|---|
| Accuracy at 3%~10% water content | Relative error absolute value not greater than 5% |
| Accuracy at 15%~25% water content | Relative error absolute value not greater than 2.5% |
| Accuracy at 35%~45% water content | Relative error absolute value not greater than 5% |
| Repeatability | Not greater than 2% (repeatability standard deviation) |
| Operating temperature | -10 C to 55 C |
| Relative humidity | 95% at 40 C, no condensation |
| Power supply | DC; recommended 12 V and 24 V; normal operation within -15% to +20% of rated voltage |
| Signal output | Digital RS232, RS485 or SDI-12; analog 1 V~5 V or 4 mA~20 mA |
| Power consumption | Working current generally not greater than 200 mA; standby current generally not greater than 1 mA |
| Vibration test | 10 Hz~150 Hz~10 Hz sweep at 2g acceleration |
| Free fall test | 30 cm vertical height |
| Reliability | Mean time between failures not less than 16000 h |
| Stable time | Not greater than 10 s for FDR, TDR and capacitance instruments |
The measurement range should preferably be 0~50% soil volumetric water content.
The instruments should work normally at temperatures from -10 C to 55 C.
The mean time between failures should be not less than 16000 h.
DC power is used, with recommended voltages of 12 V and 24 V.