Brookhaven Instruments DLS Sample Preparation Guide

This document describes how to perform particle size measurements using DLS. It also discusses how to choose the appropriate solvent and diluent, and how to prepare the sample.


Brookhaven Instruments DLS Sample Preparation Guide - cover page
Brand
Brookhaven Instruments
Category
Analyzers
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
253 KB
Published
17 November, 2023
Updated
17 November, 2023
MD5 checksum
23C97E8B7084998A3D397AF80D4A7539

About this manual

This technical note explains how to prepare samples for dynamic light scattering (DLS) measurements. It defines solvent and diluent terms and gives guidance for aqueous and non-polar liquids, dry powder dissolution, liquid sample dilution, concentration limits, colored solutions, filtering, and final measurement checks.

It recommends screening particle charge with a trace amount of salt such as 10 mM KNO3, filtering aqueous diluents, keeping samples clear to very slightly hazy, and verifying concentration with count-rate or dilution checks before recording DLS results.

What's inside

  • Nomenclature
  • Aqueous measurements
  • Diluent / Liquid
  • Dry sample dissolution
  • Liquid sample preparation
  • Concentrations
  • Colored solutions
  • Filtering the solution
  • Measurements

Frequently asked questions

Should DLS particle size measurements be made in pure de-ionized water?

No. The electrical double layer around particles can make the measured size too large by 2 to 10 nm, so a trace amount of salt is recommended to screen charge.

What salt concentration is recommended for aqueous diluents?

A concentration of 10 mM KNO3 is recommended as ideal for all concentrations of particles.

What filter pore size can be used in most cases for DLS?

A 5 micrometer filter can be used in most cases, with the largest measured size smaller than the pore size by a factor of 3.

How can the user check that the sample concentration is not too high?

The count rate should be less than 2 Mcps with intensity maximized, with a maximum measurement count rate of 500-600 kcps; a 50% dilution check can also confirm suitability.