The PhyMetrix Loop-powered Moisture analyzer is a high-precision, high-reliability moisture analyzer. It uses innovative Loop-powered technology to achieve fast, accurate moisture content measurement without the need for batteries or an external power supply. The analyzer is suitable for use in a variety of industries, including food, pharmaceuticals, chemicals, agriculture and more.
The PhyMetrix PLMa series includes the PLMa, PLMaEx, and PLMa-Probe-Ex loop-powered moisture analyzers. They measure moisture content in gases from -110°Cdp to +20°Cdp and provide a 4-20mA output with HART digital communication. The PLMa is a blind two-wire transmitter, while the PLMaEx includes a built-in sampling system and the PLMa-Probe-Ex inserts directly into pressurized pipelines.
The manual covers principles of operation, precautions, sampling techniques, mechanical and electrical installation, HART interface, troubleshooting, maintenance, and specifications. It also includes installation drawings for each model and calibration log appendices.
| Measurement range | -110°Cdp to +20°Cdp |
|---|---|
| Output signal | 4-20mA linear to selected units |
| Default scaling | 4mA = -100°Cdp, 20mA = +20°Cdp |
| Communication interface | HART (standard), Modbus RS-485 (optional) |
| PLMa process connection | 3/4"-16 parallel threads with O-ring seal |
| PLMaEx sampling system | Stainless steel 100 micron filter and flow control orifice |
| PLMa-Probe-Ex insertion | Direct insertion into live pressurized pipeline |
| PLMa-Probe-Ex fittings | 1/4" or 3/8" Swagelok tube fittings |
| Maximum sample pressure | 1000 psia |
The analyzer measures moisture content in gases from -110°Cdp to +20°Cdp.
It provides a 4-20mA output linear to the selected measurement units.
No, the PLMa is a loop-powered blind transmitter with no display.
The analyzer includes a HART interface and an optional Modbus RS-485 interface.