PhyMetrix PLMa, PLMaEx, PLMa-Probe-Ex Loop-Powered Moisture Analyzer User Manual

The PhyMetrix Loop-powered Moisture analyzer is a high-precision, high-reliability moisture analyzer. It uses innovative Loop-powered technology to achieve fast, accurate moisture content measurement without the need for batteries or an external power supply. The analyzer is suitable for use in a variety of industries, including food, pharmaceuticals, chemicals, agriculture and more.


PhyMetrix PLMa, PLMaEx, PLMa-Probe-Ex Loop-Powered Moisture Analyzer User Manual - cover page
Brand
Phymetrix
Category
Analyzer
Document type
User Manual
Language
English
Pages
29
File format
PDF
File size
1.3 MB
Published
17 November, 2023
Updated
17 November, 2023
MD5 checksum
9CA577E93E2198B0DA1B196CD05815CD

About this manual

The PhyMetrix PLMa series includes the PLMa, PLMaEx, and PLMa-Probe-Ex loop-powered moisture analyzers. They measure moisture content in gases from -110°Cdp to +20°Cdp and provide a 4-20mA output with HART digital communication. The PLMa is a blind two-wire transmitter, while the PLMaEx includes a built-in sampling system and the PLMa-Probe-Ex inserts directly into pressurized pipelines.

The manual covers principles of operation, precautions, sampling techniques, mechanical and electrical installation, HART interface, troubleshooting, maintenance, and specifications. It also includes installation drawings for each model and calibration log appendices.

What's inside

  • Introduction
  • Important Principles of Operation
  • Precautions
  • Moisture System Considerations
  • Sampling Techniques
  • Installation
  • Digital Interface - HART
  • Troubleshooting and Maintenance
  • Specifications
  • Appendices

Specifications

Measurement range-110°Cdp to +20°Cdp
Output signal4-20mA linear to selected units
Default scaling4mA = -100°Cdp, 20mA = +20°Cdp
Communication interfaceHART (standard), Modbus RS-485 (optional)
PLMa process connection3/4"-16 parallel threads with O-ring seal
PLMaEx sampling systemStainless steel 100 micron filter and flow control orifice
PLMa-Probe-Ex insertionDirect insertion into live pressurized pipeline
PLMa-Probe-Ex fittings1/4" or 3/8" Swagelok tube fittings
Maximum sample pressure1000 psia

Frequently asked questions

What is the measurement range of the PhyMetrix PLMa moisture analyzer?

The analyzer measures moisture content in gases from -110°Cdp to +20°Cdp.

What output signal does the PLMa provide?

It provides a 4-20mA output linear to the selected measurement units.

Does the PLMa have a display?

No, the PLMa is a loop-powered blind transmitter with no display.

What communication interfaces are available?

The analyzer includes a HART interface and an optional Modbus RS-485 interface.

Download PDF (1.3 MB)