IMTRON SIM-DMS Manual Strain Gauge Calibrator Product Information

The IMTRON Manual Strain Gauge Calibrator is used to simulate strain gauge sensors with bridge resistances of 120 Ω, 350 Ω or 1000 Ω. The strain gauge calibrator is used to test or calibrate strain gauge amplifiers.


IMTRON SIM-DMS Manual Strain Gauge Calibrator Product Information - cover page
Brand
Imtron
Category
Calibration Components
Document type
Product Leaflet
Language
English
Pages
1
File format
PDF
File size
93 KB
Published
17 November, 2023
Updated
17 November, 2023
MD5 checksum
37456323A45C97519772C54A6BD4814E

About this manual

The IMTRON SIM-DMS manual strain gauge calibrator is used to simulate strain gauge sensors with bridge resistances of 120 Ω, 350 Ω, or 1000 Ω. It is designed for testing and calibrating strain gauge amplifiers. Sensitivity settings are adjusted manually via high-precision switches, and polarity switching is achieved by reversing the shunt circuit.

The device offers sensitivity settings from 0 to 50 mV/V in discrete steps, with an accuracy and stability of ±0.02% for settings of 2 mV/V or higher. It operates in temperatures from 0 to 50 °C and measures 118 × 64 × 170 mm. The standard terminal is a 7-pole Lemo jack, with optional banana jacks available.

What's inside

  • Characteristics
  • Technical Data
  • Ordering Code
  • Sensitivity Settings
  • Bridge Resistances
  • Terminals
  • Accuracy
  • Stability
  • Environmental Temperature
  • Dimensions

Specifications

Sensitivity settings0; 0.25; 0.5; 1; 2; 4; 5; 8; 10; 25; 50 mV/V
Bridge resistances120, 350, 1000 Ω
Terminal7-pole Lemo jack series 2B
Accuracy±0.02% (for setting ≥2 mV/V)
Stability±0.02% (for setting ≥2 mV/V)
Environmental temperature0..50 °C
Dimensions118 × 64 × 170 mm (B × H × T)

Frequently asked questions

What bridge resistances does the IMTRON SIM-DMS calibrator support?

It supports 120 Ω, 350 Ω, and 1000 Ω bridge resistances.

What sensitivity settings are available on the manual strain gauge calibrator?

Sensitivity settings include 0, 0.25, 0.5, 1, 2, 4, 5, 8, 10, 25, and 50 mV/V.

What is the accuracy of the calibrator?

The accuracy is ±0.02% for settings of 2 mV/V or higher.

What terminal connection does the device use?

The standard terminal is a 7-pole Lemo jack series 2B.